Troubleshooting fluidics automation error on the NextSeq 500/550

Possible causes of the fluidics automation error on the NextSeq 500/550 include:

  • Communication issue between the hardware and software.

  • The waste/spent reagent container is not fully inserted.

  • A hardware component issue.

These can occur during NextSeq 500/550 control software initialization, during a run, or during a wash.

Example notification in NextSeq 500/550 control software:

Descriptions of Fluidics Automation Errors that can be observed in the control software:

  • 1: Operation home failed due to flowcell_z not being able to move to flowcell_z_home=1

  • 2: Operation home failed due to flowcell_z not being able to move to flowcell_z_home=0

  • 3: Failed to move flowcell_z to down

  • 6: Move flow cell to out failed due to flow cell home not being 0

  • 7: Failed to move flowcell_x (FlowCell Loading Mechanism)

  • 12: Failed to move flowcell_align to out

  • 20: Failed to move sipper_y to pick

  • 22: Move reagent z to down failed due to reagent tray up not being 0

  • 24: Operation home failed due to reagent_z not being able to move to reagent_tray_down=1

  • 25: Operation home failed due to reagent_z not being able to move to reagent_tray_down=0 (System.ApplicationException)

  • 26: Operation home failed due to reagent_z not being able to move to reagent_tray_down=0

  • 30: Sensor align flow not aligned to 1

  • 34: Operation home failed due to flowcell y not being able to move to flowcell_y_home=1

  • 35: Sipper_Y not being able to move

  • 36: Operation home failed due to sipper_y not being able to move to sipper_y_home=1 - fluidics

  • 39: Cannot perform pierce. Machine state is unk and is expected to be loaded. The mechanisms have been automatically reinitialized in order to return to a working state

  • 41: Cannot perform unload. Machine state is unk

  • 47: Operation up failed due to straw_z not being able to move to straws_down=0

  • 50: Operation up failed due to straw_z not being able to move to straws_up =1

  • 51: Operation up failed due to straw_z not being able to move to straws down = 0

  • 52: Operation down failed to waste presence not being 1

  • 54: Operation down failed due to straw_z not being able to move to straws_down=1.

  • 55: Operation up failed due to straw_z not being able to move to straws_down=0

  • 58: Related to not loading tray

  • 65: Sensor reagent_door_open should be 0, but it is not

  • 66: Sipper_y stopped due to safety_interlock

  • 67: Reagent_z stopped due to safety_interlock

  • 69: Operation init failed due to flowcell_align1 not being 1 (or 0)

Troubleshooting steps:

  1. First, reinsert the waste/spent reagent container (all the way to the back of the compartment).

  2. Power cycle\* NextSeq 500/550.

  1. Perform a Quick Wash.

For reference, please see Knowledge Base articles:

For any feedback or questions regarding this article (Illumina Knowledge Article #1384), contact Illumina Technical Support techsupport@illumina.com.