# Troubleshooting fluidics automation error on the NextSeq 500/550

Possible causes of the fluidics automation error on the NextSeq 500/550 include:

* Communication issue between the hardware and software.
* The waste/spent reagent container is not fully inserted.
* A hardware component issue.

These can occur during NextSeq 500/550 control software initialization, during a run, or during a wash.

Example notification in NextSeq 500/550 control software:

![](/files/7kfQ6yNFAlkfSIoEUMCh)

**Descriptions of Fluidics Automation Errors that can be observed in the control software:**

* 1: Operation home failed due to flowcell\_z not being able to move to flowcell\_z\_home=1
* 2: Operation home failed due to flowcell\_z not being able to move to flowcell\_z\_home=0
* 3: Failed to move flowcell\_z to down
* 6: Move flow cell to out failed due to flow cell home not being 0
* 7: Failed to move flowcell\_x (FlowCell Loading Mechanism)
* 12: Failed to move flowcell\_align to out
* 20: Failed to move sipper\_y to pick
* 22: Move reagent z to down failed due to reagent tray up not being 0
* 24: Operation home failed due to reagent\_z not being able to move to reagent\_tray\_down=1
* 25: Operation home failed due to reagent\_z not being able to move to reagent\_tray\_down=0 (System.ApplicationException)
* 26: Operation home failed due to reagent\_z not being able to move to reagent\_tray\_down=0
* 30: Sensor align flow not aligned to 1
* 34: Operation home failed due to flowcell y not being able to move to flowcell\_y\_home=1
* 35: Sipper\_Y not being able to move
* 36: Operation home failed due to sipper\_y not being able to move to sipper\_y\_home=1 - fluidics
* 39: Cannot perform pierce. Machine state is unk and is expected to be loaded. The mechanisms have been automatically reinitialized in order to return to a working state
* 41: Cannot perform unload. Machine state is unk
* 47: Operation up failed due to straw\_z not being able to move to straws\_down=0
* 50: Operation up failed due to straw\_z not being able to move to straws\_up =1
* 51: Operation up failed due to straw\_z not being able to move to straws down = 0
* 52: Operation down failed to waste presence not being 1
* 54: Operation down failed due to straw\_z not being able to move to straws\_down=1.
* 55: Operation up failed due to straw\_z not being able to move to straws\_down=0
* 58: Related to not loading tray
* 65: Sensor reagent\_door\_open should be 0, but it is not
* 66: Sipper\_y stopped due to safety\_interlock
* 67: Reagent\_z stopped due to safety\_interlock
* 69: Operation init failed due to flowcell\_align1 not being 1 (or 0)

**Troubleshooting steps:**

1. First, reinsert the waste/spent reagent container (all the way to the back of the compartment).
2. Power cycle\\\* NextSeq 500/550.

* If control software fails to initialize with same notification, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further guidance.

3. Perform a Quick Wash.

* If wash completes without error **or** all system check tests pass:
  * Proceed with the next run and monitor performance. See [NextSeq 550 system specifications](https://www.illumina.com/systems/sequencing-platforms/nextseq/specifications.html) to verify performance.
* If the error reoccurs **or** poor run performance is observed, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for further guidance.

For reference, please see Knowledge Base articles:

* **\***[**How to perform a NextSeq 500/550 Power Cycle**](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000001972)
* [**How to perform washes on the NextSeq 500/550**](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000003076)
* [**How to perform a NextSeq 550 system check**](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001883)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1384), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001384%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001384.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.