> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000005647.md).

# How to troubleshoot unable to locate D:\ drive or cartridge failed to unload error on iSeq 100

This issue has been observed after an iSeq 100 factory reset and/or a software re-install.

**Error message**: The system could not restore to a safe state prior to exiting, retry, if failure persists hold the button on the side of the instrument until the lights dim.

* The above error usually coincides with other errors, such as:
  * Cartridge failed to unload. Retry unloading the current cartridge
  * Unable to connect to the D:\ drive
  * Process Management cannot display runs because the hard drive D:\ is missing

**Troubleshooting steps:**

1. Check if both error messages: “Cartridge failed to unload. Retry unloading the current cartridge” and “Unable to exit application…” coincide.
2. In iSeq Control Software (ICS), navigate to Process Management and check for any errors similar to the “Unable to connect to the D:\ drive” error.
3. Check for drives to see if D:\ is present in Windows file explorer.
4. If D:\ drive is absent and instead E:\ drive is present, navigate to Disk Management by right-clicking on the windows start button, and change the drive letter from E to D.
5. Restart Windows and reopen ICS.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5647), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005647%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000005647.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.