Troubleshooting 'Lane Pump not initialized' fluidics related error on the NextSeq 500/550

Last updated
Was this helpful?
Background
This error indicates an issue with the fluidics system and can occur at anytime (eg, during initialization, washes, pre-run checks, or during sequencing runs). The error may result from a communication issue between the hardware and software, or from an issue with the syringe pump itself.
Error text:
"Error in Fluidics.FlowCellFluidics1.LanePump. Device not initialized. Command = 'Q', reply = '7'"

Troubleshooting steps:
Power cycle the NextSeq 500/550 following the steps outlined in How to Power Cycle the NextSeq 500/550
If the instrument initializes successfully, complete a Quick Wash as per instructions in How to perform washes on the NextSeq 500/550
If the wash completes without errors, set up the next run on the instrument.
If the error persists after initialization, contact Illumina Technical Support for additional guidance
For any feedback or questions regarding this article (Illumina Knowledge Article #2103), contact Illumina Technical Support techsupport@illumina.com.