> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009320.md).

# How to set up a PhiX validation run on MiSeq i100 Series

**Background**

* A PhiX validation run confirms proper hardware and software performance of the instrument. The [PhiX Control v3 Library](https://knowledge.illumina.com/library-preparation/general-library-prep/library-preparation-general-library-prep-reference_material-list/000001545.md) is a well-balanced genome with relatively equal representation of A, T, G, and C nucleotides.
* PhiX lacks an index and is not an appropriate tool for assessing Index Read or DRAGEN Secondary Analysis performance.
* Below are instructions for setting up a PhiX validation run on MiSeq i100 series sequencing platforms. The optimal loading concentration for a PhiX validation run on the MiSeq i100 series is 120 pM.

**Cloud Run Planning**

* To plan run in the Cloud using BaseSpace Sequence Hub, use the Run Planning tool in BaseSpace Sequence Hub to set up a sequencing run. For more information on BaseSpace Sequence Hub, refer to the [BaseSpace Sequence Hub support site page](https://support.illumina.com/sequencing/sequencing_software/basespace.html).

**Local Run planning**

1. Select the menu icon in the upper-left corner.
2. Select **Runs**.
3. On the Planned tab, select **Create run**.
4. Enter a run name to identify the run.
5. Enter the number of cycles performed in each read
   1. For the 100 cycle kit, use 51 cycles for Read 1 and Read 2
   2. For the 300 cycle kit, use 151 cycles for Read 1 and Read 2
   3. For the 600 cycle kit, use 301 cycles for Read 1 and Read 2
   4. For the 1000 cycle kit, use 501 cycles for Read 1 and Read 2 (the [PhiX Indexed Control](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000009867.md) is recommended for this setup due to its bigger insert size).
   5. Set number of cycles for Index 1 and Index 2 to 0 (zero) for all kits. Do not leave these fields blank.
6. Select **Next**.
7. Select the analysis application, typically **DRAGEN BCL Convert**.
8. Select the library prep and index adapter kits, typically **Not Specified**.
9. Select **Next**.
10. Input the Sample ID, typically **PhiX**.
11. Select the FASTQ file compression format, typically **gzip**.
12. Select **Next**.
13. Select **Save as Planned**.

**Start a Sequencing Run**

1. Select **Start**.
2. Sign in with local or cloud account.
3. Select the PhiX run configured in the steps above from the **Planned Runs** list.
4. Select **Review** and then review your run information.
5. Select **Load consumables**.
6. Place the new dry cartridge in the dry cartridge tray. Gently push the dry cartridge until it touches the back of the tray, so it is firmly secured.
7. Select **Next**.
8. Remove the plastic cap and load the wet cartridge.
9. Select **Close**.
10. Select **Verify Run**.
11. Confirm the run and consumables are correct, and then select **Start run**.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9320), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009320%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009320.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.