> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-troubleshooting-list/000001216.md).

# Single Read option not available with Resequencing Analysis for the TruSeq HT kit on the MiniSeq

Users attempting to resequence a run that used a TruSeq HT sample kit may find that the "Single-Read" option is not available on the MiniSeq in Local Run Manager.

Troubleshooting steps:

* Create a custom sample kit with the same definition as the TruSeq HT sample kit found in the kit Definition .tsv file (this can be downloaded from the Illumina Website under the appropriate "Product Files" Page).
* Confirm that the index strategy and read type values are All, then upload the Custom kit.

This issue has also been resolved in Local Run Manager v2.4+, which is available on MiniSeq Control Software v2 for Windows 10.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1216), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001216%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-troubleshooting-list/000001216.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.