> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000006675.md).

# Software update best practices on the iSeq 100

**General Notes**

Updating the software ensures that the system has the latest features and fixes. Software updates are bundled into a system suite, which includes the following software:

* iSeq 100 Control Software
* iSeq 100 sequencing recipes
* Universal Copy Service
* Real-Time Analysis
* Local Run Manager (framework only)
  * Although Local Run Manager is included in the system suite, the analysis modules are not. Install them separately as needed, using the sbsadmin account. Access the analysis module software from the [Local Run Manager support page](https://support.illumina.com/sequencing/sequencing_software/local-run-manager.html).

The system is configured to download software updates automatically or manually.

* Automatic updates: Updates are automatically downloaded from BaseSpace Sequence Hub to the instrument to install. This option requires an internet connection, but does not require a BaseSpace Sequence Hub account.
* Manual updates: Updates are manually downloaded from the web, saved locally or to a portable device, and installed from the saved location. This option does not require an internet connection.

**Install an Automatic Software Update**

1. Switch to the **sbsadmin** operating system (OS) account.
2. Select the control software menu, and then select **Software Update** to open the Software Update dialog box.
   1. Systems configured for automatic updates display an alert when a software update is available.
3. To check for an update, select one of the following options.
   1. Check for Update: Check for a software update.
   2. Autocheck for Updates: Check for a software update and configure the system to automatically check for future updates.
4. Select **Update** to download the new version of software.
   1. When the download is complete, the control software closes and the installation wizard appears.
   2. The control software automatically restarts. Any firmware update occurs automatically after the restart.
   3. Canceling an update is not possible after installation begins. It is only possible to cancel an update during the download step.

**Install a Manual Software Update**

1. Switch to the sbsadmin operating system account.
2. When a software update is available, download the suite installer (\*.exe) from the [iSeq 100 Sequencing System support page](https://support.illumina.com/downloads.html?filters=web%253Asystems%252Fiseq). Save the installer to a local or portable drive.
3. If the installer is saved to a portable drive, plug the drive into a USB port on the back of the instrument.
4. From the control software, select **Software Update** from the control software menu.
5. In the Software Update dialog box, expand **Install** from local or portable drive.
6. Select **Browse** to navigate to the installer.
7. Select **Update** to start installation.
   1. The control software closes and the installation wizard appears.
   2. The control software automatically restarts. Any firmware update occurs automatically after the restart.
   3. Canceling an update is not possible after installation begins. It is only possible to cancel an update during the download step.

See [iSeq 100 Product Documentation Software Updates](https://support-docs.illumina.com/IN/iSeq100/Content/IN/iSeq/SoftwareUpdates_fISQ.htm) for more information.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6675), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006675%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000006675.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.