> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000009365.md).

# Troubleshooting 'A configuration file cannot be assessed' Warnings during Pre-Run Checks on the NovaSeq X Series

**Background**

The NovaSeq X Series Control Software version v1.3 introduces a new RTA Distortion Calibration to improve run performance and accurate base calling. The values for this calibration are stored in a configuration file that is assessed during Pre-Run Checks.

If the RTA Distortion Calibration file is not configured during the Control Software v1.3 upgrade, then the configuration file will not be available and the following Warning will present during Pre-Run Checks:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-dce57b3dac6ede250d1ead9d2ad1b6c5e3786077%2Fimage1.png?alt=media)

`A configuration file cannot be assessed. This will not impact current run setup. After starting run, contact Tech Support for assistance.`

If this Warning presents, select **Continue** to proceed with the sequencing run and contact Illumina [Tech Support](https://www.illumina.com/company/contact-us.html) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9365), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009365%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000009365.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.