# Does library loading differ between MiSeq Micro and Nano flow cells?

Libraries will be denatured and diluted to the same final volume of 600 ÂµL regardless of MiSeq flow cell is used.

If a library loading concentration has been optimized using MiSeq v2 reagents with a standard flow cell, the same loading concentration should be used with a Micro or Nano flow cell. If a library loading concentration has been optimized with MiSeq v3 reagents, the loading concentration may need to be slightly lowered when using a Micro or Nano (or standard output v2) flow cell due to the lower cluster density supported with MiSeq v2 reagents.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5606), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005606%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-faq-list/000005606.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.