# Troubleshooting "Local Run Manager Error Row was updated or deleted" error on the NextSeq 500/550

Users may notice the following message appear after sequencing is complete: "Local Run Manager Error - Row was updated or deleted by another transaction (or unsaved value mapping was incorrect). Details service unavailable."

This error message indicates a communications error on the instrument and usually does not indicate and impact on run quality or proper run completion.

Once the current process on the instrument is complete, such as the post-run wash, perform a power cycle to reset the instrument's communications systems.

* For further information, search for Knowledge Base article **How to Perform a NextSeq 500/550 Power Cycle**.

If the instrument is unable to re-initialize, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for next steps.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7036), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007036%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000007036.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.