> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002543.md).

# How to stop and restart Universal Copy Service manually on the NovaSeq 6000

Universal Copy Service (UCS) copies output files from RTA3 and NovaSeq Control Software (NVCS) to the output folder throughout a run. Restarting UCS may be required if data transfer is stalled while copying to BaseSpace Sequence Hub and/or a network location.

**To restart UCS:**

* Make sure that the instrument is not running on either side.
* Close NovaSeq Control Software.
* Navigate to Task Manager and end task for UCS process (called ucs.exe or universal-copy-service): right-click on the process and choose "End Task."
* Restart UCS either by:
  * Re-starting the NovaSeq Control Software. Restarting control software automatically reboots Universal Copy Service.
  * Or by navigating into C:\ProgramFiles\Illumina\UniversalCopyService and double-clicking ucs.exe or universal-copy-service.exe.
    * If restarting from C:\ProgramFiles\Illumina\UniversalCopyService, a black window appears to show UCS progress.
* **Note**: Upload may be slightly delayed. To assess the progress of upload to BaseSpace, view the file size increase in the run summary.
* If the issue with data transfer persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for the next steps.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2543), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002543%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002543.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.