# GenomeStudio Methylation Module why are some beta values blank even though there are intensities reported for them?

In GenomeStudio Methylation projects, some Beta values are blank even though there are intensity values reported in the Signal A and Signal B columns. The reason usually is that the number of beads used to calculate an intensity falls below the threshold of 3 for either the red channel or the green channel. If either the intensity from signal A or signal B is calculated from less than 3 beads, then GenomeStudio will not calculate or report a Beta value.

In order to confirm this, the user should add the subcolumns "Avg\_NBEADS\_A" and "Avg\_NBEADS\_B" to the Sample Methylation Profile,\
![](/files/97j5eT4xVcAgfGVqUbCp)

then sort the Avg Beta column A-Z for a single sample. Avg\_NBEADS reports the number of beads used in calculating the reported signal.\
![](/files/GZbEIPT1hYlou41L758u)

The blank Beta values will appear at the top. For these markers, the user will find one or both of Avg\_NBEADS\_A or Avg\_NBEADS\_B will have fallen to less than 3. Note that there is no way to bypass this threshold in GenomeStudio, so if the user wishes to calculate a beta value for such assays, they will have to do so manually outside of GenomeStudio.

ImproveCtrlD\
\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3003), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003003%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-troubleshooting-list/000003003.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.