# What effect does prolonged flow cell staging time have on clustering for the NovaSeq 6000?

NovaSeq 6000 Standard Workflow:

Prolonged staging time between denaturation and the start of the sequencing run can cause denatured libraries to reanneal, leading to reduced clustering efficiency.

NovaSeq 6000 Xp Workflow:

Prolonged flow cell staging time between ExAmp master mix combination with pooled libraries and the start of the sequencing run can cause increased % occupancy and lower % pass filter. As a general rule, the maximum recommended flow cell staging time is 30 minutes.

![](/files/tkhyhqzZCEVyhX1kn5hy)

For more information, see the technical note, [The NovaSeq Xp workflow provides flexibility and control without sacrificing data quality or yield](https://support.illumina.com/content/dam/illumina-marketing/documents/products/technotes/novaseq-xp-workflow-tech-note-770-2018-014.pdf).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7323), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007323%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-faq-list/000007323.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.