What effect does prolonged flow cell staging time have on clustering for the NovaSeq 6000?
Last updated
Was this helpful?
Last updated
Was this helpful?
NovaSeq 6000 Standard Workflow:
Prolonged staging time between denaturation and the start of the sequencing run can cause denatured libraries to reanneal, leading to reduced clustering efficiency.
NovaSeq 6000 Xp Workflow:
Prolonged flow cell staging time between ExAmp master mix combination with pooled libraries and the start of the sequencing run can cause increased % occupancy and lower % pass filter. As a general rule, the maximum recommended flow cell staging time is 30 minutes.