> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007479.md).

# How to perform a Maintenance Wash on the NovaSeq X Series Instruments

**Background**

The NovaSeq X Series instruments perform an automated wash after every successful sequencing run. As part of normal instrument maintenance, an additional maintenance wash is required every **14 days** or whenever an automatic post-run wash fails to complete.

The maintenance wash requires:

* Wash Reagent cartridge
* Used Sequencing Buffer cartridge
* Either a Wash flow cell or a used 8-lane sequencing flow cell that has not been removed from the instrument
* Empty used Small and Large Spent Reagents Bottle

**Important Notes**

*For the NovaSeq X Control Software v1.3 and above:*

* The Wash setup is only available from the **Start Menu** in the NovaSeq X Control Software, the Wash option has been removed from Settings.
* Illumina recommends performing Maintenance Washes with the used 8-Lane flow cell that has not been removed from the instrument whenever possible.
* The Wash Flow Cell usage count is now enforced and will prevent using a Wash Flow Cell with over 20 uses
  * The Control Software will present a warning when use count reaches 15
  * For assistance with replacing the Wash Flow Cell, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html).

*For the NovaSeq X Control Software v1.2.2 and below:*

* The wash setup is available from either the Control Software **Settings Menu** or the **Start Menu**.
* The Wash Flow Cell usage count is NOT enforced, though the Software will add to the use counter for that Wash Flow Cell.
  * Illumina recommends performing Maintenance Washes with the used 8-Lane flow cell that has not been removed from the instrument whenever possible.

**Preparing the Wash Solution Buffer**

1. Add 400 mL laboratory-grade water to a 500 ml centrifuge bottle.
2. Add 0.2 mL 100% Tween-20 to the laboratory-grade water to prepare the 0.05% Tween-20 wash solution.
   * Using a freshly prepared dilution of Tween-20 limits the introduction of contaminants into the fluidics system.
3. Invert solution several times to mix thoroughly.
4. Add 380 mL of 0.05% Tween-20 wash solution to the large wash solution well in the **wash reagent cartridge**.
5. Combine the following volumes in a 50 mL centrifuge tube to prepare 49 mL of 0.12% reagent grade NaOCl:
   1. 1.2 mL 5% reagent grade NaOCl.
   2. 48 mL Deionized (DI) water.
6. Invert solution several times to mix thoroughly.
7. Add 49 mL 0.12% NaOCl solution to the bleach well on the **wash reagent cartridge**.

**Setting up the Maintenance Wash (Run Time: \~3 hours)**

*For NovaSeq X Control Software v1.3 and above:*

1. From the **Start** Menu in the NovaSeq X Series Control Software home screen, select **Wash**.

*For the NovaSeq X Control Software v1.2.2 and below:*

1. From the NovaSeq X Series Control Software home screen, select the instrument icon to open the global navigation menu.
2. Select **Settings**, then select **Wash**.
3. Select the checkboxes for the instrument side(s) to perform the Maintenance Wash on.
   1. If performing a Maintenance Wash on one side, the other side will not be accessible during the Maintenance Wash.
4. If a used 8-Lane sequencing flow cell that is still in place from the last sequencing run, the instrument will automatically load the used flow cells and the user can skip to Step 5.
   1. Users can utilize an 8-lane used sequencing flow cell to perform a maintenance wash only if it remains loaded on the flow cell stage after the previous run.
   2. Once the sequencing flow cell is unloaded from the flow cell stage it can no longer be used for a maintenance wash.
   3. If the used 8-lane flow cell has been removed, proceed to step 3.
5. Select the checkbox **Load wash flow cells** to replace the used sequencing flow cell with aWash flow celland then select the **Load wash flow cells** button.
   * After selecting, the display monitor raises and the flow cell door opens. The flow cell light indicates the instrument side which the wash is being performed.
6. Once the flow cell door has opened, place the **wash flow cells** in the flow cell stage and select **Close flow cell door**.
   * Wash Flow Cells are currently limited to 20 uses or its expiration date, whichever comes first.
7. On the Wash screen, select **Load wash reagent cartridges**.
   * The instrument doors unlock automatically, and the system displays information on loading the wash reagent and buffer cartridges.
8. Remove the used sequencing reagent cartridge, then load the Wash Reagent Cartridge into the instrument in the right bay of the reagent drawer.
   * A **previously-used buffer cartridge** is required to start a maintenance wash and should be left in place.
9. Empty the used reagent bottles (small and large) then returned the used reagent bottles into the used reagent drawers.
   * Dispose of spent reagents according to local EH\&S guidelines.
10. Close the reagent drawers, then close the instrument doors.
11. Select **Verify wash cartridges**.
12. After all wash consumables are verified, select **Start wash**.
13. After starting the wash, the instrument doors automatically lock. The doors unlock after the wash completes and the instrument returns to the Start screen.

For most sequencing run issues, the instrument will attempt to perform the automatic post-run wash, so a maintenance wash will not be required for many run-stopping issues.

For detailed instructions, see the **Perform a Maintenance Wash** section of the [NovaSeq X Plus Product Documentation](https://support.illumina.com/sequencing/sequencing_instruments/novaseq-x-novaseq-x-plus/documentation.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7479), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007479%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007479.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
