> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/iscan/microarray-iscan-troubleshooting-list/000003861.md).

# How to check for corrupt DMAPs before scanning BeadChips using the iScan

DMAP files may transfer incorrectly to the network during downloading and file integrity can be compromised. The iScan can be configured to check the DMAP file integrity at the start of each scan.

**Note:** Enabling this feature increases the amount of time before the iScan begins performing the scan.

**To verify DMAP file integrity**:

1. From the menu in the upper left corner of the screen, select **Tools** > **Options**.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-1d1a1d8022968a0373e2d67db2114e74a7d019d8%2Fimage1.jpg?alt=media\&token=b665bd35-1434-457d-9e14-54146674768a)
2. In the Options dialog box, select the **General** tab.
3. In the Processing section, select the **Enable Corrupt DMAP Check** checkbox, and then select **OK**.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-7e11b8c3d37d453262151f66271bd6bab7f5f477%2Fimage2.jpg?alt=media\&token=3f72118d-35f4-4838-bb2a-55051b73f8f6)

For additional information, see the [iScan System Product Documentation](https://support.illumina.com/downloads/iscan-system-product-documentation.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #3861), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000003861%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/iscan/microarray-iscan-troubleshooting-list/000003861.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.