# Troubleshooting validation errors displaying when opening "module create run" page on the MiniSeq

This validation error may appear if the installed module that the user is attempting to set up a run with is not compatible with the installed version of Local Run Manager (LRM).

To resolve this issue:

* Navigate to C:\Illumina\LocalRunManager\Modules\<Module Name>\<Module Version>\UI\\
* Confirm only 2 files present:
  * 1 modules. \\\*.jsfile
  * 1 modules-templates. \\\*.jsIf
* Update the module to the most recent or compatible version. Modules are available on the [Illumina Support Page for Local Run Manager](https://support.illumina.com/sequencing/sequencing_software/local-run-manager/downloads.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1221), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001221%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |