> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000009386.md).

# Troubleshooting Stalled Secondary Analyses Onboard the MiSeq i100 Series

The MiSeq i100 Series instrument allows onboard DRAGEN Analysis of sequencing run data. If this process stalls or is not able to complete, the Run Details page shows the Secondary Analysis as 'In progress' and the File transfer status as 'Not started'. See the following mitigations below if this issue is identified.

**Too many external storage configurations**

On the MiSeq i100 Series instruments, it is possible to configure more external storage mounts than are supported. The maximum supported number of external configurations is two.

If three or more external storage configurations are in place, it is possible the Secondary Analysis will appear to stall, and never start Secondary Analysis Data Transfer. The image below is an example of what the status of the run may look like in this scenario.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-31953368addf2944cc3dbc6044e0c370f5682e1b%2Fimage1.png?alt=media)

If the system is configured with three or more external storage mounts, follow the steps below to resolve the issue.

**Resolution actions**

1. Cancel the ongoing analysis:
   1. From the Control Software Home screen, select the **dropdown menu** in the top-left corner, then select **Runs**.
   2. Within the Runs user interface (UI), select the **Active** tab.
   3. Under the Run Name column, select the **sequencing run** to open the Run Details UI.
   4. From within the Run Details UI, select scroll down to the Secondary Analysis header.
   5. Select **Cancel Analysis**.
2. From the Control Software Home screen, select the **dropdown menu** in the top-left corner, then select **Settings**, then **External Storage**.
3. Remove one External Storage location to leave a maximum of two configurations in place.
   1. See [MiSeq i100 Product Documentation - External Storage](https://support-docs.illumina.com/IN/MiSeqi100Series/Content/IN/MiSeqi100/SysOutFolder.htm) for additional details.
4. Requeue Secondary Analysis:
   1. From the Control Software Home screen, select the dropdown menu in the top-left corner, then select **Runs**.
   2. Within the Runs UI, select the **Completed** tab.
   3. Under the Run Name column, select the **sequencing run** to open the Run Details UI.
   4. Scroll down to the Secondary Analysis header, then select **Requeue Analysis**.
   5. Follow the prompts to requeue the analysis using the Run Planning tool or from a sample sheet.
   6. Use the dropdown menu under Output folder location to **select the intended output folder** on the configured External Storage location.
   7. Input the Reanalysis Reason.
   8. Select **Requeue Analysis**.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9386), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009386%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-troubleshooting-list/000009386.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.