Troubleshooting install package integrity check failed during NextSeq 1000/2000 upgrade to v1.7
Background
The NextSeq 1000/2000 Control Software upgrade requires admin access. If the upgrade is being attempted from an account that does not have admin privileges, an error will occur:
"Install package integrity check failed, Software updated was unsuccessful. Re-launch NextSeq 100/200 Control Software to retry."

To successfully update the control software complete the Troubleshooting actions included below.
Troubleshooting actions
Select the Power button in the upper right corner of the screen.
Navigate to the listed user and select logout.
Login as ilmnadmin or other admin account.
Once logged in with an account with proper access, the control software upgrade will be able to proceed as expected.
For any feedback or questions regarding this article (Illumina Knowledge Article #8890), contact Illumina Technical Support techsupport@illumina.com.