# How to troubleshoot errors during the pre run Wet Checks on the NovaSeq X/X Plus

**Background**

The Pre-Run Checks are a series of tests that help ensure the instrument is ready to perform a sequencing run. They consist of Software, Dry, and Wet Check stages. The Wet Checks will pierce the buffer cartridge and flow buffer through the flow cell to confirm that the fluidics circuit is performing as expected.

If the instrument is unable to verify the fluidics circuit is working as expected, the Control Software will present a notification indicating the Wet Checks were unsuccessful.

*Recovering the Reagent* *Cartridge*

If a dual-sided run was setup and the Pre-Run Checks were unsuccessful on only one side, see the following steps to recover the Reagent Cartridge with the Lyo Insert and Library Strip tube from the affected Side.

1. On the Pre-Run Checks screen, select **Cancel Side \[A/B]** to cancel the affected run and allow the other Side to proceed.
2. Fill ALL wells of a **Wash Reagent Cartridge** with **DI water**.
3. Once the Run has started on the successful side, select the the **instrument icon**, then select **Settings**, then select **Unlock Doors**.
4. From the Unlock Doors user interface (UI), select the checkbox for the affected Side and then select **Unlock**.
   1. The side currently sequencing will be greyed out and cannot be selected.
5. Once the door is open, remove the **Reagent Cartridge** then load the filled **Wash Reagent Cartridge** on the affected side.
   1. The **Reagent Cartridge** can be transferred to another NovaSeq X Series instrument for immediate use, stored at 4° C for up to 72 hours, or re-frozen once at -20° C for longer term storage.
6. Leave the spent **Buffer Cartridge** in place on the affected side.
7. Close the reagent drawer and the instrument door, then select **Lock Doors**.
8. Once the Run completes on the successful side, see the *Troubleshooting Actions* below.

*Troubleshooting Actions*

If this issue presents during the Pre-Run Checks, collect the following information and files from the instrument:

1. Remove the flow cell and visually confirm if flow cell lanes appear wet or dry
2. Inspect the flow cell gasket and the instrument flow cell holder/manifold for any liquid or precipitate
3. Collect the run setup logs from the following directory based on the instrument side that presented the error from the following directory(ies):
   1. /usr/local/illumina/runs/setup-A or ../setup-B

Once this information is collected, contact [Illumina Tech Support](https://www.illumina.com/company/contact-us.html) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7951), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007951%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000007951.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
