Post-Run Automatic Purge for NextSeq 500/550

Background

The purge option appears when the run stops prior to completion (eg, due to a clustering failure). Select Yes to initiate purging of the reagent.

Estimated time for automatic purge

  • The time for automatic post-run purge completion depends on the length of the sequence run and the type of kit used. This can take between 15 minutes and several hours.

    • The purge duration depends on how much left-over reagent remains in the reagent cartridge.

    • After the purge, a post-run wash will automatically start (it takes 90 minutes).

  • Users can expect the purge + post run wash to take between 2-5 hours.

Additional notes

  • Users can choose to opt out of the purge; however, a manual post-run wash must be performed prior to setting up the next run.

  • There is an option to set up an automatic reagent purge after each run, to purge unused reagents from the reagent cartridge to the spent reagents container automatically after each run. For further information, see the NextSeq 550 System Guide.