> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009362.md).

# Troubleshooting Inaccurate Estimated Completion Time on the MiSeq i100 Series

**Background**

The MiSeq i100 Series instrument uses data from previous sequencing runs to provide an accurate estimated time to completion for ongoing sequencing runs. Generally the estimated time is accurate to within an hour, and becomes increasingly accurate as the instrument collects data from completed sequencing runs.

If the estimated time to completion is off by more than an hour after completing more than 5 sequencing runs, see the following Troubleshooting Actions.

**Troubleshooting Actions**

1. Log into the Control Software as an Administrator user.
2. Select the **dropdown menu** in the top-left corner, then select **Settings**.
3. Select **Time settings**.
4. Use the **Time zone** dropdown menu to select the correct local time zone.
5. \[Optional] Input the local Network Time Protocol (NTP) address.
   1. This is only required if local network settings prevent the instrument from accessing the internet and a local NTP server has been configured. Consult the local IT team for additional assistance.
6. Select **Save**.

If the estimated time to completion is not corrected after the next sequencing run, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9362), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009362%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009362.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.