> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000002353.md).

# Troubleshooting "Y home sensor never tripped" error on MiSeq

The MiSeq may give the error "Y stage encountered homing error when MiSeq was doing “Home All” function" or “Y home sensor never tripped" during MiSeq Control Software (MCS) initialization or during a run. This error will likely cause an ongoing run or initialization to fail.

**Troubleshooting steps**:

1. Perform a post-run wash if the run stops prematurely or after the run is completed.
2. Power cycle the MiSeq following the steps outlined in [How to Power Cycle the MiSeq](https://knowledge.illumina.com/instrumentation/miseq-ruo/instrumentation-miseq-ruo-reference_material-list/000001894.md).
3. Re-start the MCS and, if no issues with re-initialization, perform the **Y Stage test five times.** Instructions on how to perform system checks are outlined in [How to perform a System Check on the MiSeq.](https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-reference_material-list/000001777.md)
   * Make sure that a used flow cell is loaded on the instrument (eg, the same flow cell used for the wash).
   * Select Y Stage test.
     * For MCS v4.0 and above: From the Main Menu, select System Check.
     * For MCS v3.1 and below: From the Welcome Screen, select Manage Instrument, then select System Check.
   * Follow the on-screen prompts to start the test, and then repeat the test four more times (for a total of five times).
4. If all Y Stage tests pass, continue with normal sequencing.
5. If the system fails to initialize or any of the five replicates of the Y Stage test fail, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2353), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002353%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000002353.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.