# Troubleshooting a Failed to Get Temperature error on the NextSeq 500/550

The following warning is typically a transient communication issue between the NextSeq Control Software and a temperature sensor and does not affect run performance.

***"@@@---Failed to get temperature."***\
![](/files/surhSXzCezXemFdj9dAU)

**Troubleshooting steps:**

1. If run performance is normal, power cycle instrument the instrument after the run completes and continue with the next run if the system initializes without issue.
2. If run performance is affected, power cycle instrument and perform system check tests. If all tests pass, then proceed with next sequencing run.
3. If the warning re-occurs or the system check shows a failure, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/).

For reference, search Knowledge Base articles:

* **How to Perform a NextSeq 500/550 Power Cycle**
* **How to perform a NextSeq 500/550 system check**

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1382), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001382%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001382.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.