> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009239.md).

# Troubleshooting 'Failed to start First Time Setup' Errors on the MiSeq i100 Series Instruments

**Background**

The First Time Setup workflow assists with setting up an Administrator account, networking, and performing a System Health Check. If the Control Software is not able to proceed with starting the System Health Checks, the following error message displays:

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-e14044791eb7a174b17553945c1b911e3a2ccab7%2Fimage1.png?alt=media)

**Figure 1**: Critical System Error: Failed to start First Time Setup: Analysis is ongoing. System Checks cannot be started.

This issue generally presents when the DRAGEN Manager service is attempting to perform the DRAGEN license check and is not getting a response, typically due to network security restrictions. If this issue presents, see the following instructions.

**Troubleshooting Actions**

1. Allow the instrument to remain idle for **minimum of 1 hour** on the error screen.
2. Select **Restart application** to restart the Control Software. Do not power cycle the instrument.
3. After initialization, the instrument will restart **First Time Setup**. Navigate through the screens, confirming that the information was correctly saved from the first attempt.
4. Once the First Time Setup is complete, proceed to starting the **System Health Checks**.

If the issue persists, please contact [Illumina Tech Support](https://www.illumina.com/company/contact-us.html) for further assistance.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #9239), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000009239%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-troubleshooting-list/000009239.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.