> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001479.md).

# How to degas the Te-Flow

**The proper technique for degassing the Te-flow is as follows.**

* Set water circulator to 44°C in preparation for XStain.
* When the temperature reaches \~ 30°C, separate the heat exchanger (black piece) from the reagent pan (white base) (A).
* Lift the heat exchanger upright and away from you (tubing at the bottom) (B) and Turn 90° counterclockwise (C).
* Turn back horizontal (A). Repeat (B) > (C) > (A) 3 times for a total of 4 rotations or until air/bubbles are no longer audible.
* Using Kimwipes dampened with lab-grade water, clean all surfaces between the heat exchanger and reagent pan as indicated by the blue arrows shown in D and E to remove reagent build-up. Discard Kimwipes with formamide waste.
* Place Te-flow back on the reagent pan and ensure it is flush using the two guide pins in the reagent pan.

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-4ca223c4c9c6b01fb6d9197883a40e11d2e83db8%2Fimage1.png?alt=media)

This information is also available in video form. Search for Knowledge Base article **How to degas the Te-flow Video.**

For further information, search for Knowledge Base Article **Infinium Array: Best Practices Support Webinar Video.**

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1479), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001479%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/general/microarray-general-reference_material-list/000001479.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.