For the complete documentation index, see llms.txt. This page is also available as Markdown.

Troubleshooting camera, cluster, and focus errors at cycle 1 on the MiniSeq

The following notification types indicate that the MiniSeq is unable to detect clusters at cycle 1. Cycle 1 errors can occur due to instrument, consumable, or library (design/custom primers/quantification/denaturation) causes.

Potential error messages include:* Camera 1 disabled

  • Failure to Detect Clusters

  • Failed to generate focus model for the top surface

Note 1: A pop-up in the Control Software appears asking if the user would like to save the flow cell. Select No, as the flow cell cannot be re-used.

Note 2: The instrument gives an option to purge reagents after the run ends. This option will proceed with purging consumables, then performing the automatic post-run wash. The purging process can take a few hours (especially for a cycle 1 error because most of the reagents are still in the reagent cartridge). The purge is optional and can be bypassed if desired.

Troubleshooting steps:

  1. Acknowledge the error message.

  • The flow cell cannot be saved.

  • Illumina does not require the reagent purge.

  1. Perform the post-run wash.

  2. Power cycle the instrument.

  1. Perform the System Check with a used flow cell.

  1. If the System Check passes, the failure may be due to an issue with the library or reagents.

  2. To investigate the most common issues:

If there are no apparent issues with the library or reagents, repeat the run with a 20% PhiX control spike in. This will act as a positive control for clustering for the next run to determine if an underlying library issue caused the previous error.

If an error occurs at cycle 1 of the next run, Contact Illumina Technical Support at techsupport@illumina.com for further assistance.

For any feedback or questions regarding this article (Illumina Knowledge Article #1198), contact Illumina Technical Support techsupport@illumina.com.

Last updated

Was this helpful?