Troubleshooting "Ebox0 Fan is not operating as expected" warnings on the MiniSeq

Background

The "Ebox0 Fan is not operating as expected" warning during runs on the MiniSeq does not impact sequencing quality. This warning may indicate a communications issue on the instrument or the need for maintenance on the fan itself.

Troubleshooting steps:

  1. Contact Illumina Technical Support to document the issue.

  2. Allow the run to complete and perform the post-run wash.

  3. Power cycle the MiniSeq.

    1. For further information, see Illumina Knowledge Article How to Power Cycle the MiniSeq.

    2. If the Control Software fails to initialize, contact Illumina Technical Support.

  4. If the instrument initializes without errors, perform the Thermal Fan test through the MiniSeq System Checks.

    1. Navigate to Home screen in MiniSeq Control Software (MNCS).

    2. Select Manage Instrument --> System Check

    3. Follow the instructions to prepare the instrument for the check as indicated.

    4. Ensure only the following checks are selected:

      1. Thermal: Fans

  5. If the system checks pass, this indicates the issue is related to a communications issue on the instrument and has been resolved. Proceed with the next run.

  6. If the system checks fail, contact Illumina Technical Support for further assistance.