Pre Run Check Troubleshooting on the MiniSeq

If the MiniSeq pre-ruck checks do not pass, follow the guidance below for the check or checks that did not pass.

General/reagent checks

  • Doors closed.

    • Make sure doors closed.

    • Power cycle\* if issue persists.

  • Consumables loaded.

    • Ensure consumables are properly loaded.

    • Power cycle\* if issue persists

  • Required Software.

    • Verify that the proper software/module is installed.

  • Instrument and network disk space checks (the control software will provide a warning if the free storage is below 100Gb).

    • Clear disk space from the D:\ drive.

    • Clear disk space from the network output location, if needed.

  • Network connection.

    • Confirms network status and physical network connection.

    • Check read/write permissions, availability of network location, and Ethernet plug.

Temperature

  • Temperature ramp: checks ramp of flow cell heater plate.

    • Power cycle\* and retry the pre-run checks.

  • \*\*Temperature sensors.

    • Power cycle\* and retry the pre-run checks.

  • \*\*Fans

    • Power cycle\* and retry the pre-run checks.

Imaging System

  • \*\*Imaging limits: sends stage to sensor limits.

    • Power cycle\* and retry the pre-run checks.

  • \*\*Z step-and-settle: checks the ability of z-stage to move and stop moving in a set time.

    • Power cycle\* and retry the pre-run checks.

  • Bit Error rate: checks camera circuits working.

    • Power cycle\* and retry the pre-run checks.

  • \*\*Flow cell registration: checks if the flow cell is in the correct imaging position.

    • Clean and re-seat flow cell and check gaskets.

    • Power cycle\* and retry the pre-run checks. If the checks do not pass, try a new flow cell.

Reagent Delivery

  • \*\*Valve response: tests the accuracy of valve and pump syringe rate movements.

    • Power cycle\* and retry the pre-run checks.

  • Pump: moves pump syringe to min and max positions.

    • Power cycle\* and retry the pre-run checks.

Note: Pre-run checks will not check flow rates, so always visually inspect the flow cell gaskets prior to loading.

\*For further information, search for Knowledge Base article How to Power Cycle the MiniSeq.

\\ These checks are also performed during the system check process.

If any pre-run check fails after a full power cycle, contact Illumina Technical Support.

For any feedback or questions regarding this article (Illumina Knowledge Article #1214), contact Illumina Technical Support techsupport@illumina.com.

Last updated

© 2023 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. Trademark information: illumina.com/company/legal.html. Privacy policy: illumina.com/company/legal/privacy.html