Pre-run check troubleshooting on the MiniSeq
If the MiniSeq pre-run checks do not pass, follow the guidance below for the checks that did not pass:
Reagent checks:
Doors closed.
Make sure doors are closed.
Power cycle* if issue persists.
Consumables loaded.
Ensure consumables are properly loaded.
Power cycle* if issue persists.
Required Software.
Verify that the proper software/module is installed.
Instrument and network disk space checks - the control software will provide a warning if the free storage is below 100Gb.
Clear disk space from the D:\drive.
Clear disk space from the network output location, if needed.
Network connection.
Confirms network status and physical network connection.
Check read/write permissions, availability of network location, and Ethernet plug.
Temperature:
Temperature ramp: checks function of flow cell heater plate.
Power cycle* and retry the pre-run checks.
Temperature sensors**.
Power cycle* and retry the pre-run checks.
Fans**.
Power cycle* and retry the pre-run checks.
Imaging System:
Imaging limits**: moves stage to sensor limits.
Power cycle* and retry the pre-run checks.
Z step-and-settle**: checks the ability of z-stage to move and stop moving in a set time.
Power cycle* and retry the pre-run checks.
Bit Error rate: checks camera circuits.
Power cycle* and retry the pre-run checks.
Flow cell registration**: checks if the flow cell is in the correct imaging position.
Clean and re-seat flow cell and check gaskets.
Power cycle* and retry the pre-run checks. If the checks do not pass, try a new flow cell.
Reagent Delivery:
Valve response**: tests the accuracy of valve and pump syringe rate movements.
Power cycle* and retry the pre-run checks.
Pump: moves pump syringe to min and max positions.
Power cycle* and retry the pre-run checks.
Note:
Pre-run checks will not check flow rates, so always visually inspect the flow cell gaskets prior to loading.
*For further information, search for Knowledge Base article How to Power Cycle the MiniSeq. **These checks are also performed during the system check process.
If any pre-run check fails after a full power cycle, contact Illumina Technical Support.