Troubleshooting "An unexpected error occurred during sequencing" Errors on the MiSeq i100 Series

Background

Issues during a sequencing run that the Control Software is not able to categorize will stop the run and present the following generic error message in the Notifications menu:

"An unexpected error occurred during sequencing."

If this issue presents during a sequencing run, please see the following Troubleshooting Actions.

Troubleshooting Actions

  1. Select Eject Cartridge, to remove the dry and wet sequencing cartridges from the instrument.

  2. Run the instrument System Health Checks using the Test Cartridges provided with the instrument.

  3. Once the System Health Checks are complete, select Clear Checks and then proceed to the Control Software home screen.

  4. Open the menu and select Settings. Then select Export Logs

    1. Select the Standard option to bundle the System Health Check results.

    2. Select the affected Sequencing Run from the list when setting up the Export Logs.

    3. See Knowledge Article How to Export Logs on the MiSeq i100 Series for more information.

  5. Once these actions are completed, contact Illumina Tech Support for next steps.

For any feedback or questions regarding this article (Illumina Knowledge Article #9264), contact Illumina Technical Support techsupport@illumina.com.