> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000005499.md).

# Troubleshooting NovaSeq 6000 Reagent Chiller outside acceptable range warning

**Background**

The NovaSeq 6000 reagent chiller remains at room temperature when the instrument is idle. At the start of a sequencing run, the instrument will begin cooling the reagent chiller compartment (RCA) on the instrument and will target a temperature of 7-8°C and maintain that temperature for the duration of the run. Issues with maintaining the appropriate reagent chiller temperature can cause run quality issues.

Issues during the run, if present, will often present the following **warning message**:

21-06-08 13:28:28.818 -05 0208 WRN: The chiller returned a temperature of 23.0289993286133 at the beginning of sequencing, which is outside the acceptable range of \[4, 14] -- SequenceCommand.cs, CheckChillerTemperature, 2729

**Troubleshooting steps**:

* Check if the ambient temperature is significantly higher than expected.
  * Maintain a lab temperature of 19°C to 25°C (22°C ± 3°C). For further information, see [NovaSeq Sequencing System Site Prep Guide](https://support.illumina.com/downloads/novaseq-site-prep-guide-1000000019360.html).
* Collect Run Cycle logs: C:\ProgramData\*\Illumina\NovaSeq\NovaSeqTemp\[run folder]\Logs (zipped)
  * After collecting the logs, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) to determine the next steps.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #5499), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000005499%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-troubleshooting-list/000005499.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.