> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000007019.md).

# Troubleshooting Black Chromium Screen on the iSeq 100

**Background**

When users open Chromium on the iSeq 100 they may find the window is entirely black. Sometimes users may be able to type in the address bar, but the window itself will not change even when refreshed. This may result from a issue on the instrument or a side-effect of hardware acceleration in Google Chromium.

**To troubleshoot this issue:**

1. Power Cycle the instrument as described in the [How to Power Cycle the iSeq 100](https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-reference_material-list/000002289.md). Reopen Chromium once the instrument re-initializes.
2. Turn off Hardware Acceleration in Chromium as follows:
   1. Create a Google Chromium shortcut on the Windows Operating System (OS) Desktop.
   2. Right-click the shortcut, and select **Properties**.
   3. Select **Shortcuts** tab.
   4. In the Target window, add the following text: `"[space]-disable-gpu"`
      1. The Target path should show `chrome.exe" -disable-gpu"`.
   5. **Save the** changes.
   6. Open Chrome and select **Settings** > **Show advanced settings**.
   7. Disable the **Use hardware acceleration when available** option.

If these steps do not re-establish Chromium's normal function, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7019), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007019%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/iseq-100/instrumentation-iseq-100-troubleshooting-list/000007019.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.