> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000006309.md).

# Calculating Percent Passing Filter for Patterned and Nonpatterned Flow Cells

**Introduction**\
The latest innovation in flow cell technology is the development of the patterned flow cell. Five Illumina sequencing platforms currently take advantage of this advanced technology: the iSeq 100, MiSeq i100 Series, NextSeq 1000/2000, NovaSeq 6000 and NovaSeq X Series System. Patterned flow cells consist of a nano well substrate with billions of ordered wells (Figure 1A). Compared to nonpatterned flow cells (Figure 1B), the uniform cluster sizes enable optimal spacing and increased cluster density. Although the ordered spacing of patterned flow cells enables significantly higher cluster density, patterned flow cells also report comparatively lower percent passing filter (%PF) values\* due to differences in the %PF calculation method. This technical note outlines these differences and describes how they lead to lower %PF metrics for patterned flow cells.

**Percent Passing Filter with Nonpatterned Flow Cells**

Real-Time Analysis (RTA) software proceeds through several stages including image analysis, template generation, base calling, passing filter calculations, and quality scoring. Template generation occurs during cycles 1-5 of Read 1 and defines the position of each cluster in a tile. The template is used as a reference for registration and intensity extraction during subsequent sequencing cycles. With nonpatterned flow cells, dim or low-quality clusters are removed from the raw cluster count during template generation (Figure 2). This effectively acts as a prefiltration step, removing clusters unlikely to pass filter in the first 25 sequencing cycles and yielding relatively high %PF values.

**Percent Passing Filter with Patterned Flow Cells**

With patterned flow cells, the calculation of %PF is different because there is no template generation step—fixed cluster locations eliminate the need for template generation. Because template generation and the associated preliminary filtration steps are not applied, empty wells or clusters that may be dim, low quality, or polyclonal are included in the raw cluster count, which leads to lower %PF values. Although the percent passing filter metric will be much lower with patterned flow cells, it will not impact performance or data quality.

**Summary**

Due to the differences in the %PF calculation methods, patterned flow cells have artificially lower %PF metrics compared to nonpatterned flow cells. With patterned flow cells, there is no template generation or preliminary filtration step during image analysis, which results in lower %PF values. Although the %PF values are lower, the uniform feature sizes and optimal spacing enable significantly increased cluster density for patterned flow cells.

**\*** The %PF calculations involve the application of a chastity filter to each cluster. Chastity is defined as the ratio of the brightest base intensity divided by the sum of the brightest and second brightest base intensities. Clusters “pass filter” if no more than 1 base call has a chastity value below 0.6 in the first 25 cycles. This filtration process removes the least reliable clusters from the image analysis results.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-ea03d688f437f2d4712b48ee8726bddb34b55b69%2Fimage1.png?alt=media\&token=2405492e-ae74-4f50-bde1-b48be8f98f3d)

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-5d13439180edab78e833586a77243e56e132b9d1%2Fimage2.png?alt=media\&token=c79f300c-d915-4efb-8f24-63660cc902df)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #6309), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000006309%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/general/instrumentation-general-reference_material-list/000006309.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.