Skip to content
For the complete documentation index, see llms.txt. This page is also available as Markdown.

How to set up a PhiX validation run on the NextSeq 500/550 in manual mode

Background

Guidance

  • Below are instructions for setting up a PhiX validation run for the NextSeq 500/550 system in Manual mode.

  • Manual mode allows you to setup the run without using Local Run Manager or BaseSpace.

  1. Open the NextSeq 500/550 Control Software (NCS).

  2. Select Experiment.

  3. Select Sequence.

  4. Select Manual.

  5. Edit the run parameters in Run Setup; an example paired-end run is shown below.

Note: For the run, set both Index 1 and Index 2 to zero. Example above is when using a 150 cycles kit. Some validation runs aim to test the overall capacity (stress test), while others analyze clustering efficiency. Adjust read length according to the purpose of the validation run and kit type.

  1. Follow the prompts on the screen to load the flow cell.

  2. Follow the prompts on the screen to discard spent reagents, reload the spent reagents tray, and load a new buffer cartridge.

  3. Load the reagent cartridge.

  4. Review the run configuration.

  5. Next, perform the instrument self-check, after which the PhiX validation run begins sequencing.

Additional Information

For more information about preparing PhiX for the NextSeq 500/550 sequencing platform, see the following documents:

For any feedback or questions regarding this article (Illumina Knowledge Article #1532), contact Illumina Technical Support techsupport@illumina.com.

Last updated

Was this helpful?