> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000008271.md).

# How to troubleshoot "The instrument encountered a thermal coolant error" on the NovaSeq X Series

**Background**

The NovaSeq X Series of instruments utilize several coolant loops to ensure normal thermal operation of the instrument. Issues with any of these Thermal Loops will typically present as Critical system errors that prevent or stop sequencing runs to ensure the instrument does not become damaged. These issues typically present with the following error message:

"Critical System Error: The instrument encountered a thermal coolant error. Please shutdown the instrument and contact Illumina Technical Support".

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-30bcd659e3db7449e7fcf6c03d89927e248b57bf%2Fimage1.png?alt=media\&token=cb1ae78a-0b5d-4817-abfb-2dc798cf5dea)

**Troubleshooting Actions**

1. Contrary to the error statement, **do not** immediately power cycle the instrument.
2. Collect the following Logs depending on error context:
   1. If during a sequencing run, collect the following files:
      1. The Logs.zip file from the `/usr/local/illumina/runs/[RunID]/Logs` directory
      2. `The entire InstrumentAnalyticsLogs folder in the /usr/local/illumina/runs/[RunID] directory`
   2. If during initialization or instrument idle, collect the following files:
      1. The entire novaseq-x-series-control-software folder in the `/usr/local/illumina/logs` directory
      2. The entire non\_run folder in the `/usr/local/illumina/performance_analytics_data/` directory
3. Contact [Illumina Tech Support](https://www.illumina.com/company/contact-us.html) for next steps to troubleshoot this issue.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #8271), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000008271%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-troubleshooting-list/000008271.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.