> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001542.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# Spiking custom primers into the Illumina sequencing primers

Libraries prepared with custom sequencing primers require optimization to perform successfully on Illumina sequencing platforms.

Combining (or spiking) custom primers into the Illumina primers is necessary when including PhiX or other standard Illumina libraries with the custom library type. Illumina cannot guarantee the performance or compatibility of custom primers, and it is up to each lab to validate primers for sequencing on a specific platform.

When using custom sequencing primers, it is critical to know the recommended platform-dependent loading concentration, total volume, loading positions, and primer annealing temperatures (see [Considerations when migrating non-Illumina libraries between sequencing platforms](https://knowledge.illumina.com/library-preparation/custom-third-party-library-prep/library-preparation-custom-third-party-library-prep-reference_material-list/000001478.md) for more information). This article provides information for spiking custom primers into the Illumina primers.

**Spiking custom primers into Illumina sequencing primers**

* The cartridge position, total volumes, and final concentration of custom primers for each platform are provided in the tables below.
* Calculate the volume of the custom primer to add to the Illumina primer cartridge position based on the final concentration of the custom primer in the cartridge\*.
* After spiking in the custom primer, adjust with an appropriate pipette to half of the final volume, and at the cartridge position gently pipette up and down five times to mix thoroughly\*\*.
* For MiSeq, MiniSeq, NextSeq 500/550 and NovaSeq 6000 platforms, do **not** check the "custom primer" box position in the sample sheet or during the run setup.

\* To calculate how much custom primer to spike into the well, use the (C1)(V1) = (C2)(V2) equation where:\
C1 = the concentration of your custom primer (when using high concentration primer stock, such as 100 µM, the small additional volume to the final volume is negligible).\
V1 = solve for the volume of the custom primer to be spiked in\
C2 = the recommended custom primer final concentration from the chart below\
V2 = total volume of Illumina primer in the charts below

Example: (100 µM) (V1) = (0.5 µM) (680 µl)\
V1 = 3.4 µl

\*\* MiSeq, NextSeq 500/550 and MiniSeq cartridges: to increase successful measurement of the Illumina sequencing primers, the addition and mixing of custom primers, it is recommended to use slimmer tips or larger volumes of lower stock concentration custom primers.

**Important Note**: The guidelines below are based on the current primer volumes. To ensure accuracy, measure total primer volumes in the cartridge with a pipette before proceeding with setup.

**iSeq 100**

With the release of iSeq Control Software version 3.0, it is possible to sequence with custom primers on the iSeq 100. Due to the construction of the iSeq 100 cartridge, it is not possible to spike custom primers directly into the pre-loaded Illumina primers.

In lieu of spiking custom primers directly into reagent cartridge wells, the NextSeq 1000/2000 custom primer kit configurations are compatible with the iSeq 100 and should be used. Custom primers should be added to the appropriate Illumina primers, then loaded into the iSeq 100 reagent cartridge using the custom primer workflow. For available kit configurations and their part numbers, refer to the [NextSeq 1000/2000 Product Documentation](https://support-docs.illumina.com/IN/NextSeq10002000/Content/IN/FrontPages/NextSeq10002000.htm).

**Important Note:** The NextSeq 1000/2000 Custom Primers Guide provides primer addition guidelines with volumes specific to the NextSeq 1000/2000. See below for iSeq 100-specific guidance for adding custom primers.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-f07db8338b3f7bc4a555e34f65ef434a4b984ef8%2Fimage1.jpg?alt=media\&token=bb1df546-0bc0-4f53-b39a-f443311bdf22)

**MiniSeq**\
Because of the unique dual-indexing workflow on the MiniSeq, combinations of custom primers and Illumina primers are limited. Be aware of the custom primer dependencies outlined [here](http://support.illumina.com/downloads/miniseq-custom-primers-guide-1000000002700.html).

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a2ee7f03a3f62d65773103b4f5c721823e593481%2Fimage2.jpg?alt=media\&token=dd491c40-af9e-400f-9f95-5c6753755dff)

**MiSeq**

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-f0c4e46a8f20e161e772e8ab8138c9e334fe6c70%2Fimage3.jpg?alt=media\&token=102bec07-eb15-48f9-8761-50d7d3ffaa8f)

\* There is no option for a custom Index 2 (i5) primer since the template uses the grafted P5 primer on the surface of the flow cell.

**MiSeq i100 Series**

For information on spiking custom primers into a MiSeq i100 run, refer to [Spiking custom primers into the Illumina sequencing primers for MiSeq i100 Series](https://knowledge.illumina.com/instrumentation/miseq-i100-series/instrumentation-miseq-i100-series-reference_material-list/000009551.md)

**NextSeq 500/550**\
Because of the unique dual-indexing workflow on the NextSeq 500/550, combinations of custom primers and Illumina primers are limited. Be aware of the custom primer dependencies outlined [here](https://support.illumina.com/downloads/nextseq-500-custom-primers-guide-15057456.html).

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-86cecd279f5d39159e456d870b71a4ceb629f17b%2Fimage4.jpg?alt=media\&token=5487a1a3-c437-44e6-8d59-b60c16d4ab7a)

**NextSeq 1000/2000**

Due to the construction of the NextSeq 1000/2000 cartridge, it is not possible to load custom primers into the Illumina primers.

If Illumina primers are needed to sequence PhiX during a run, a NextSeq 1000/2000 Custom Primer kit is required. Refer to the [NextSeq 1000/2000 Product Documentation](https://support.illumina.com/sequencing/sequencing_instruments/nextseq-1000-2000/documentation.html) for instructions.

**NovaSeq 6000 v1.5 Consumables**\
Because of the primer design and dual-indexing workflow on the NovaSeq 6000, combinations of custom primers and Illumina primers are limited. Be aware of the custom primer dependencies outlined [here](https://support.illumina.com/downloads/novaseq-custom-primers-guide-1000000022266.html).

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-cfe469887f7c7bffcb6227cb21958faef25fbcd3%2Fimage5.png?alt=media)

**NovaSeq X Series**

The NovaSeq X Series uses **two** primer solutions that combine primers for Read 1 and 2, and Index 1 and 2. If custom primers are only required for specified Reads, spike-in the concentrated primers into the appropriate wells for the Reads that require custom primers. When spiking in custom primers, **no changes** to the run set up are needed and a custom recipe is not required.

**Table** provides the well positions, Illumina primer volume, and the recommended custom primer (CP) volume to add in microliters (µl) **assuming a starting concentration of 100 micromolar (µM) for the custom primer stock that will be used**. Adjust volumes as needed for different custom primer stock concentrations.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-5d530c1dad30d065a18d9ed81e930b3286816064%2Fimage6.jpg?alt=media)

**Using Custom Primers in Place of Illumina Primers**

If only the custom primer is needed, custom primers can be used in place of the Illumina primers. For instructions on how to set up a run using only custom primers, see the following guides:

* [MiniSeq System Custom Primers Guide](http://support.illumina.com/downloads/miniseq-custom-primers-guide-1000000002700.html)
* [MiSeq System Custom Primers Guide](http://support.illumina.com/downloads/using_customprimers_miseq_15041638.html)
* [NextSeq 500/550 System Custom Primers Guide](http://support.illumina.com/downloads/nextseq-500-custom-primers-guide-15057456.html)
* [NextSeq 1000/2000 Product Documentation - Custom Primers](https://support-docs.illumina.com/IN/NextSeq10002000/Content/IN/NextSeq2000_1000/CustomPrimers_Intro_fNS_m2000.htm)
* [NovaSeq Series Custom Primers Guide](https://support.illumina.com/downloads/novaseq-custom-primers-guide-1000000022266.html)
* [iSeq 100 Sequencing System - Custom Primers](https://support-docs.illumina.com/IN/iSeq100/Content/IN/iSeq/CustomPrimers-Intro.htm#CustPrimers-Intro)
* [NovaSeq X Series Product Documentation - Custom Primers](https://support-docs.illumina.com/IN/NovaSeqX/Content/IN/NovaSeqX/CustomPrimers.htm)

For more information on the indexing strategies for each instrument, refer to the [Indexed Sequencing Overview Guide](https://support.illumina.com/downloads/indexed-sequencing-overview-15057455.html).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1542), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001542%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000001542.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
