> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002215.md).

# Reagent Cartridge Layout for NovaSeq 6000

All reagent cartridges contain reagent-adjacent empty wells for the post-run wash.

***Cluster cartridge***

![](/files/633idK6UslmErED3qsoV)

* **Every other row in the cartridge is a WASH row. So Well 1 on the cartridge is the wash well for the reagent in well 5 (custom primer 1), etc.**
* A denaturation reagent in position #30 contains formamide, which is an organic amide and reproductive toxin.
* To facilitate safe disposal of any unused reagent after the sequencing run, this reservoir is removable.
* Do **not** stack cluster cartridges, as this can dislodge formamide position.

  ![](/files/Ttr6b5tHp4noOGsGS535)
* Position 17: 0.24% NaOCl
* Positions 5-7: custom primer positions
  * If using, specify during run setup, use HT1 to dilute
  * Position 5: Custom read 1 primer
  * Position 6: Custom read 2 primer
  * Position 7: Custom Index 1 primer and Custom Index 2 (for v1.5 kits)
* Positions 24, 13, 23: Illumina primer positions
* Position 24: Illumina read 1 primer
* Position 13: Illumina read 2 primer
* Position 23: Illumina index 1 and index 2 primers

***SBS cartridge***

![](/files/6pHA0Dfwgjmz3I9i3rpR)

![](/files/gTpeeaixHdu02Jy7Lp2w)

***Buffer cartridge***

![](/files/ViQIizeg9rCDdNG0m6TX)

* Weighs up to 15 lbs (6.8 kg).
* Wash buffers are pumped through lines to wash after each reagent pull or imaging cycle.
* Recyclable plastic (only cartridge that can be recycled).

  ![](/files/TKoiBqfhycTAC1qm3cKs)

For further information, see Knowledge Base [NovaSeq 6000 flow cell types and their compatibility with NovaSeq Control Software versions](https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002418) article.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2215), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002215%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter, and the optional `goal` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-6000/instrumentation-novaseq-6000-reference_material-list/000002215.md?ask=<question>&goal=<endgoal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is optional and describes the broader end goal you are ultimately trying to accomplish on behalf of the user. GitBook uses it to tailor the answer towards what is most useful for that goal.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.