> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/microarray/human-methylation/microarray-human-methylation-reference_material-list/000001524.md).

# Which Bisulfite Conversion Kit should be used for Infinium Methylation Arrays?

Prior to performing the Infinium Methylation Assays, all DNA samples must undergo bisulfite conversion. The efficiency and accuracy of this step are critical to overall assay performance. In particular, adherence to Illumina‑recommended protocols is essential to for optimal results.

**Recommended Bisulfite Conversion Kits for Infinium Methylation Assays**

Illumina recommends the following bisulfite conversion kits from Zymo Research, which have been validated for use with Infinium Methylation assays.

**Table 1**.Validated and recommended EZ DNA / EZ-96 DNA Methylation Kits for use with Infinium Methylation Assays

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-6e95623b1459640c5ae47628ea0a725f33efd67a%2Fimage1.jpg?alt=media)

* Approximate processing time: \~ 1.5 days
* Critical Workflow Requirements when using these kits: Follow the modified overnight/cycling incubation protocol as described in:
  * Instruction Manual - [EZ DNA Methylation Kit Catalog Nos. D5001 & D5002](https://files.zymoresearch.com/protocols/_d5001_d5002_ez_dna_methylationga_o_kit.pdf) (Appendix: Bisulfite Conversion and PCR Optimization, page 10, Step 6. Alternative Incubation Conditions When Using the Illumina Infinium Methylation Assay).
* It should be noted that the Zymo EZ DNA Methylation kits (D5001, D5002, D5004) have not been validated and are not supported for use with Infinium EX Methylation Assay. Refer to [Infinium EX Methylation DNA input guidelines and compatible bisulfite conversion kits](/microarray/human-methylation/microarray-human-methylation-reference_material-list/000008604.md) for additional information.

**Table 2**.Validated and recommended EZ-96 DNA Methylation Lightning MagPrep Kits (uses magnetic beads for purification) for use with Infinium Methylation Assays.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-f224fb372910cc301e6fb916ba7d0e78c7951d35%2Fimage2.png?alt=media)

* Approximate processing time: \~ 3 hours
* \*User provides 96-deep well microplates and PCR plates
* \*\*The EZ-96 DNA Methylation-Lightning Automation Kit is equivalent to the EZ-96 DNA Methylation-Lightning MagPrep Kit but contains buffer and reagent volumes needed for procedures using liquid-handling automation.
* **Note**: For questions regarding liquid-handling automation scripts and set-up, contact Zymo Research support at <automation@zymoresearch.com>.

**Bisulfite Conversion Kits** **that are not recommended for use with Infinium Methylation Assays**

The following bisulfite conversion kits from Zymo Research have not been validated and are not compatible with Infinium Methylation Assays.

**Table 3**.Bisulfite Conversion Kits that have not been validated and are not recommended for use with Infinium Methylation Assays.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-fc5c8d3462571aaa423c0308c079e13d77fe5bd7%2Fimage3.png?alt=media)

* **Note**:No other bisulfite conversion kits have been validated by Illumina for use with Infinium Methylation Assays.

**Additional Resources:**

* Technical Note on EZ-96 DNA Methylation-Lightning Automation Kit validation: [Automated bisulfite conversion for Infinium Methylation BeadChips](https://www.illumina.com/content/dam/illumina/gcs/assembled-assets/marketing-literature/automated-bisulfite-infinium-methylation-tech-note-m-gl-00144/automated-bisulfite-Infinium-methylation-tech-note-m-gl-00144.pdf)
* Zymo: [Bisulfite Conversion for Illumina Methylation Arrays: Troubleshooting & Best Practice Guidelines](https://www.zymoresearch.com/blogs/blog/bisulfite-conversion-for-illumina-methylation-arrays)

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                   |
| :---------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1524), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001524%20-%20Microarray%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/microarray/human-methylation/microarray-human-methylation-reference_material-list/000001524.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
