Troubleshooting MiSeqOverride.cfg Errors on the MiSeq

Background

At the start of a run the MiSeq references the MiSeqOverride.cfg file to determine calibration location. This file provides information specific for the system. During preventative maintenance or service visits, Field Service Engineers (FSEs) may make minor updates to this file. If a syntax error is introduced, the MiSeq may fail to accurately parse the MiSeqOverride.cfg and stop the next run.

  • "Config file 'C:/Miseq Config/MiSeqOverride.cfg, section '[Focus Params]' defines 'CoarseFocusCenterZ=-0.0.169263'. which is an invalid float"

Troubleshooting

The specific section and element defined may vary, and this error requires consultation with an Illumina Technical Support representative to resolve. If this warning appears:

  1. Locate the MiSeqOverride.cfg Configuration File on the instrument at based on the Windows Operating System (OS) version:

    1. Windows 7 file location: C:\MiSeq Config

    2. Windows 10 file location: C:\ProgramData\Illumina\MiSeq

  2. Email Illumina Technical Support with a picture of the error message and attach the MiSeqOverride.cfg file to your email.

A member of Illumina Technical Support will follow up to assist in troubleshooting.

For any feedback or questions regarding this article (Illumina Knowledge Article #3299), contact Illumina Technical Support techsupport@illumina.com.