For the complete documentation index, see llms.txt. This page is also available as Markdown.

What to do if insufficient wash solution was used for a manual wash on the NextSeq 500/550 or MiniSeq

If there is not enough wash solution in the wash tray during a manual wash, this can introduce air into the system which can negatively impact subsequent sequencing run performance.

To solve this, Illumina recommends performing 2-3 manual washes on the instrument to flush the fluidics lines and remove any air bubbles. See instructions on how to perform manual washes in this Knowledge Base article How to perform washes on the NextSeq 500/550 or Washes on the MiniSeq.

For any feedback or questions regarding this article (Illumina Knowledge Article #3047), contact Illumina Technical Support techsupport@illumina.com.

Last updated

Was this helpful?