Troubleshooting failed MiSeq system checks
Note: For how to perform a MiSeq system check, see Knowledge Base article How to perform a System Check on the MiSeq.
The MiSeq system checks are comprised of four tests: motion, optics, thermal ramp, and fluidics.
The tests listed under the Illumina Use Only column include the Tip/Tilt and Full Optics tests. These tests require a beaded flow cell to complete successfully. Therefore, Illumina Service Engineers carry a beaded flow cell for performing these tests.
For failed tests, power cycle the instrument, perform a maintenance wash with water and repeat the system check with an alternate flow cell. Contact Illumina Technical Support for any repeated failures or further questions.
For any feedback or questions regarding this article (Illumina Knowledge Article #7050), contact Illumina Technical Support techsupport@illumina.com.
Last updated