# Troubleshooting RFID Not Recognized Error on the MiSeq

MiSeq reagents are equipped with radio-frequency identification (RFID) tags to enable accurate consumable tracking. Independent tracking of the flow cell, reagent cartridge, and PR2 bottle occurs at the beginning of the sequencing run, when sensors at these three positions read and write to the RFID. If the RFID fails to be detected, a run can be started using a bypass code created by the MiSeq Self-Service tool available through your MyIllumina account.

To generate the code:

1. Log in to [MyIllumina](https://dashboard.my.illumina.com/).
2. Open **More** on the menu, expand **My Tools**, and select **MiSeq Self-Service**:\
   ![](/files/Hnvhbh0hNIbAzL4LRQ3g)
3. Enter the serial number of the instrument (M#####).
4. From the Type of Override Code drop-down list, select **RFID Override**.
5. Select **Get Code** to generate the code.\
   ![](/files/6Zt9kmt61HfFXI7oAswW)

The bypass code is entered directly into MiSeq Control Software and is valid for 7 days.

Follow the prompts to enter the barcode and part number for the consumable. The location of this information on the reagent label is outlined below. The barcode number is highlighted in blue, and the part number is in yellow.

![](/files/0gLjHQN4lElpKIHQoNJw)

Multiple RFID failures for the same consumable type can indicate of a faulty sensor.

For persistent issues, email Illumina Technical Support at <techsupport@illumina.com>.

For further information, search for Knowledge Base article **How to generate an RFID bypass code in MyIllumina Video.**

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1459), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001459%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/miseq/instrumentation-miseq-troubleshooting-list/000001459.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.
