> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000007990.md).

# How to fix inaccurate or inverted touch screen on the NextSeq 500/550

Following the NextSeq 500/550 Windows 10 upgrade, Hard Drive swap or LCD Screen replacement, the touch screen may respond to inputs in an unexpected manner. Touch inputs may be inverted, for example, selections at the bottom right hand corner of screen results in a click at the top left of screen or inaccurate screen calibration.

To resolve this:

1. From the Windows **Start** menu, search for "PenMount Control Panel"
   1. Open the **PenMount Control Panel**
2. Select **PenMount 6000 USB** from the main window
3. Select **Configure**
4. In the new window, open the **Calibrate** tab, then select **Standard Calibration**
5. Follow directions as prompted
6. When the control window reappears, select **OK**, then **OK** again in the next screen to exit
7. Confirm the touch screen is now responding as expected
8. If issue persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/united-states/technical-support) for assistance

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7990), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007990%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000007990.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.