How to fix inaccurate or inverted touch screen on the NextSeq 500/550

Following the NextSeq 500/550 Windows 10 upgrade, Hard Drive swap or LCD Screen replacement, the touch screen may respond to inputs in an unexpected manner. Touch inputs may be inverted, for example, selections at the bottom right hand corner of screen results in a click at the top left of screen or inaccurate screen calibration.

To resolve this:

  1. From the Windows Start menu, search for "PenMount Control Panel"

    1. Open the PenMount Control Panel

  2. Select PenMount 6000 USB from the main window

  3. Select Configure

  4. In the new window, open the Calibrate tab, then select Standard Calibration

  5. Follow directions as prompted

  6. When the control window reappears, select OK, then OK again in the next screen to exit

  7. Confirm the touch screen is now responding as expected

  8. If issue persists, contact Illumina Technical Support for assistance

For any feedback or questions regarding this article (Illumina Knowledge Article #7990), contact Illumina Technical Support techsupport@illumina.com.