# Troubleshooting Cannot Update Metrics errors on the NextSeq 500/550

This following error usually does not impact run performance or analysis. No adverse impact is expected for sequencing metrics, primary analysis, or secondary analysis in Local Run Manager.

**Error**: "*Local Run Manager Error - Cannot update metrics during state 'PrimaryAnalysisCompleted'. Originally caused by: Conflict*"

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-a95962ccaf02745423021af4449a2a845db36675%2Fimage1.png?alt=media\&token=c6303e99-96bc-4bca-bfec-c1759794efaf)

**Troubleshooting steps:**

1. Confirm the Local Run Manager Analysis has completed.
2. Power cycle the instrument.
3. If the error persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/).

For further information, see for Knowledge Base article [How to Perform a NextSeq 500/550 Power Cycle](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000001972).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2137), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002137%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |