# Troubleshooting Cannot Update Metrics errors on the NextSeq 500/550

This following error usually does not impact run performance or analysis. No adverse impact is expected for sequencing metrics, primary analysis, or secondary analysis in Local Run Manager.

**Error**: "*Local Run Manager Error - Cannot update metrics during state 'PrimaryAnalysisCompleted'. Originally caused by: Conflict*"

![](/files/eORnPXLvJ6apvtHMBcse)

**Troubleshooting steps:**

1. Confirm the Local Run Manager Analysis has completed.
2. Power cycle the instrument.
3. If the error persists, contact [Illumina Technical Support](https://www.illumina.com/company/contact-us.html#/).

For further information, see for Knowledge Base article [How to Perform a NextSeq 500/550 Power Cycle](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-reference_material-list/000001972).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2137), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002137%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions: Querying This Documentation

If you need additional information that is not directly available in this page, you can query the documentation dynamically by asking a question.

Perform an HTTP GET request on the current page URL with the `ask` query parameter:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000002137.md?ask=<question>
```

The question should be specific, self-contained, and written in natural language.
The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.