> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/software/general/software-general-reference_material-list/000001782.md). Agents can also search and read this documentation through its [MCP server](https://knowledge.illumina.com/~gitbook/mcp).

# How to requeue analysis of a run with MiSeq Reporter

MiSeq Reporter performs secondary analysis of data generated on the MiSeq instrument. The run data can be re-analyzed multiple times locally by requeueing analysis with MiSeq Reporter. Some common reasons for requeueing analysis include changing the analysis workflow or correcting index sequences, sample IDs, or samples names in the sample sheet. The instructions to edit the sample sheet correctly and requeue analysis with MiSeq Reporter, either using the web interface or from the run folder, are provided below. Always make sure that a sequencing run is not in progress when requeuing analysis with MiSeq Reporter.

**How to edit the sample sheet**

1. Navigate to the MiSeq Analysis run folder on the D drive. For example:
   1. D:\Illumina\MiSeq Analysis\141007\_M001\_0016\_000000000-AB8A8
2. To edit the SampleSheet.csv file, open in it a text editor such as Notepad. If you use a spreadsheet editor, make sure to save the sample sheet as a comma separated value (.csv) file. Make sure to maintain the format of the sample sheet.
3. If you create a new sample sheet, keep a copy of the original sample sheet by renaming the SampleSheet.csv file to SampleSheet\_original.csv. Place the new sample sheet in the MiSeq Analysis run folder and name it SampleSheet.csv.

**Requeue analysis of a run using the MiSeq Reporter web interface**

1. To launch the MiSeq Reporter web interface, open the Internet Explorer web browser and type [http://localhost:8042](http://localhost:8042/) in the address bar.
2. From the MiSeq Reporter web interface, select the Analyses tab on the top left.
3. Locate the run from the list of available runs in the Analyses tab, click on the checkbox next to the run name, then select **Requeue**.

During analysis, a status bar with elapsed time appears on the Analyses Info tab. If analysis needs to be stopped prior to completion, select the Stop analysis icon next to the status bar on the Analysis Info tab.

**Requeue without using the MiSeq Reporter web interface**

1. If the run was sent to BaseSpace Sequence Hub, but has not been analyzed locally, the run will not be listed in the Analyses section of the MiSeq Reporter web interface. In these situations, the analysis must be requeued from the run folder. Once the analysis is requeued, the run will appear in the MiSeq Reporter web interface. The following steps outline how to requeue analysis from the run folder.
2. Navigate to the MiSeq Analysis run folder on the D drive. For example:
   1. D:\Illumina\MiSeq Analysis\141007\_M001\_0016\_000000000-AB8A8
3. Delete the file called queuedforanalysis.txt (or QueuedForAnalysis.txt) in the run folder. Deleting this file from the run folder triggers MiSeq Reporter to start analysis. This file will reappear (as QueuedForAnalysis.txt) with current date and time stamp, indicating that MiSeq Reporter has started analyzing the run.

**Files for troubleshooting**

MiSeq Reporter Error.txt and MiSeq Reporter Log.txt are found in D:\Illumina\MiSeqAnalysis

To view the status of the analysis, open the Internet Explorer web browser and type [http://localhost:8042](http://localhost:8042/) in the address bar to launch the MiSeq Reporter web interface. Select the Analyses tab on the top left. The run that was just requeued will be at the bottom of the list. Select the run to see the status bar for the analysis.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                 |
| :-------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1782), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001782%20-%20Software%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/software/general/software-general-reference_material-list/000001782.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.