> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001883.md).

# How to perform a NextSeq 500/550 system check

**Important Note:** For the system checks launched from NextSeq Control Software (NCS) 4.0 and newer, the NextSeq Service Software (NSS) will present a login screen. This requires any Local Run Manager (LRM) account credentials.

**Performing the System Checks on the NextSeq 500/550**

1. Navigate to the **Home** screen in NextSeq Control Software (NCS).
2. Select **Manage Instrument** > **System Check**, then select **yes** to service software opening.
3. If prompted, input the Local Run Manager (LRM) username and password.
4. When prompted by the instrument, load the following consumables:
   1. A used high output flow cell (a high output flow cell is strongly recommended, as mid output flow cells can lead to false flow rate failures).
   2. An empty spent reagents tray.
   3. A wash buffer tray with **150 ml** of lab-grade water in the center reservoir.
   4. An **empty** wash reagent tray.
5. The tests will be pre-selected. Do not select unchecked tests, as the unchecked tests can only be performed by Illumina personnel.\
   ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-485656544b781e7afcebf0de9fdaa8d528b327bb%2Fimage1.png?alt=media)
6. Select **Next** to start the System Check. The System Checks will take about 90 minutes to complete.
   1. Passing system checks are indicated with green checkmarks and a PASS record for the test.
7. The results of the tests are saved to `D:\Illumina\SystemCheckResults\&lt;Test Name&gt;\&lt;Date of Test&gt;. Collect this folder and provide it to Illumina Support teams for further review.`
   1. A screenshot of the results summary screen is also helpful for Illumina Support.

**Note**: The new NextSeq 550 wash wash cartridge has a label listing the minimum (150 ml) and maximum allowable volumes. The previously recommended volume of 120 ml as listed in the [NextSeq 550 system guide](https://support-docs.illumina.com/IN/NextSeq_550-500/Content/IN/FrontPages/NextSeq-550-500.htm) is safe to use instead of the 150 mL listed above.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #1883), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000001883%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/nextseq-500-550/instrumentation-nextseq-500-550-troubleshooting-list/000001883.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.