Troubleshooting "Compensator Move and/or Sensor Failed" errors on the MiSeq

Users may encounter a compensator error during a MiSeq run, system check, or Control Software initialization. This error indicates that the compensator on the instrument, which allows the cameras to change focus between the top and bottom of the flow cell, is not functioning properly. This can result from a communication error on the instrument or may indicate a hardware issue on the instrument.

Error message: Compensator Move and/or Sensor Failed on retry #3 /3: Moving to OutOfPath, StartCurrent =...

Troubleshooting Steps:

  1. Perform the post-run wash.

  2. Power cycle the MiSeq. For reference, see How to Power Cycle the MiSeq or How To Power Cycle the MiSeq Video.

  3. Restart the MiSeq Control Software (MCS). If no issues with re-initialization, perform the Compensator test five times.

    1. Start the system check with the used flow cell loaded on the instrument (alternatively, the previous wash flow cell can be used). See the following Knowledge Base article for more information on How to perform a System Check on the MiSeq.

    2. Select Compensator test.

      1. MCS v4.0 and above: From Main Menu and then select System Check.

      2. MCS v3.1 and below: From Welcome Screen, select Manage Instrument, and then select System Check.

    3. Follow the prompts on the screen to start the test, and repeat the test four more times (five times total).

  4. If all Compensator test replicates pass, continue with normal sequencing.

  5. If the system fails to initialize or any of the five replicates of the Compensator test fail, contact Illumina Technical Support for further assistance.

For any feedback or questions regarding this article (Illumina Knowledge Article #8020), contact Illumina Technical Support techsupport@illumina.com.