Troubleshooting a Flow Cell Door error on the NextSeq 500/550
Last updated
Last updated
© 2023 Illumina, Inc. All rights reserved. All trademarks are the property of Illumina, Inc. or their respective owners. Trademark information: illumina.com/company/legal.html. Privacy policy: illumina.com/company/legal/privacy.html
The following error can appear during initialization or when setting up a sequencing run or wash. This can be a communication error, especially if the instrument had not been power cycled in a while.
Error message: "Exception generated from IxCmdIllumina.Spark.EmbeddedCommandSet.ix_dlowcell_door_open : An error has occurred while executing command Originally caused by: An error has occurred while executing the command"
Troubleshooting steps:
Power cycle the NextSeq 500/550.
If instrument initializes without errors, the door can be tested as follows:
Navigate to Manage Instrument and select Unload Consumables.
Proceed with the wash setup.
If the error does not re-appear, the instrument is cleared for continued use.
If the error is recurring, contact Illumina Technical Support.