> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000002344.md).

# Understanding unique dual indexes (UDI) and associated library prep kits

Along with best practices, using the unique dual indexing strategy will make sure that libraries sequence and demultiplex with the highest accuracy across all Illumina sequencing platforms. When preparing libraries for multiplexing, Illumina recommends using unique dual indexing (UDI) whenever possible for the most accurate demultiplexing.

**What to know about Unique Dual Indexing**:

* Unique dual indexing is a sequencing strategy that has distinct, unrelated index sequences for each of the i5 and i7 index reads.
* Unique dual indexing is a known mitigation for filtering index-hopped reads seen in downstream analyses. Misassigned reads will be flagged as "undetermined reads" and can be excluded from analysis. For more information about index hopping refer to [Effects of Index Misassignment on Multiplexing and Downstream Analysis](https://www.illumina.com/content/dam/illumina-marketing/documents/products/whitepapers/index-hopping-white-paper-770-2017-004.pdf). Additional best practices to mitigate index hopping can be found [here](https://www.illumina.com/science/education/minimizing-index-hopping.html).
* Unique dual indexes, or non-redundant indexes, are used in the indexing step without any changes to the library preparation workflow.
* With combinatorial dual indexing, every i5 and i7 is shared among other samples on the same plate with indexes repeating down rows and across columns. The combinations are unique though the individual indexes are not.
* In contrast to combinatorial dual indexing, unique dual indexing has completely unique indexes (eg, 96 unique i5s and 96 unique i7s per 96-well plate).

  ![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-00b3b29c9f0da2e89fea3792fae389344e364f25%2Fimage1.png?alt=media\&token=a7c25ce8-c959-4a40-9932-ef70db04a0c2)

**Figure 1:** With combinatorial dual indexing, every i5 and i7 is shared among other samples on the same plate with indexes repeating across rows and down columns; the combinations are unique though the individual indexes are not. In contrast to combinatorial dual indexing, unique dual indexing has completely unique indexes (ie, 96 unique i5s and 96 unique i7s per 96-well plate).

**The following UDI kits are available for various Illumina workflows:**

**TruSeq Unique Dual Index kits**

| Product Name                                | Catalog Number | Indexes, Samples                 |
| ------------------------------------------- | -------------- | -------------------------------- |
| IDT for Illumina - TruSeq DNA UD Indexes v2 | 20040870       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina - TruSeq RNA UD Indexes v2 | 20040871       | <p>96 Indexes,<br>96 Samples</p> |

* TruSeq DNA UDIs are compatible with [TruSeq DNA PCR-Free](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/truseq-dna-pcr-free.html) and [TruSeq DNA Nano](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/truseq-nano-dna.html)
* TruSeq RNA UDIs are compatible with [TruSeq Stranded mRNA](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/truseq-stranded-mrna.html) and [TruSeq Stranded Total RNA](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/truseq-stranded-total-rna.html)

**Illumina DNA/RNA Unique Dual Index, Tagmentation kits**

| Product Name                                    | Catalog Number | Indexes, Samples                 |
| ----------------------------------------------- | -------------- | -------------------------------- |
| Illumina DNA/RNA UD Indexes Set A, Tagmentation | 20091654       | <p>96 Indexes,<br>96 Samples</p> |
| Illumina DNA/RNA UD Indexes Set B, Tagmentation | 20091656       | <p>96 Indexes,<br>96 Samples</p> |
| Illumina DNA/RNA UD Indexes Set C, Tagmentation | 20091658       | <p>96 Indexes,<br>96 Samples</p> |
| Illumina DNA/RNA UD Indexes Set D, Tagmentation | 20091660       | <p>96 Indexes,<br>96 Samples</p> |

* All four index sets are compatible with [Nextera XT](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/nextera-xt-dna.html), [Illumina DNA Prep](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/nextera-dna-flex.html)3, [Illumina DNA Prep with Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/nextera-flex-enrichment.html)4, [Illumina DNA Prep with Exome 2.5 Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/dna-prep-exome-enrichment.html), [Illumina RNA Prep with Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/rna-prep-enrichment.html) and [Illumina DNA PCR-Free](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/dna-pcr-free-prep.html).

**Illumina RNA Unique Dual Indexes, Ligation kits**

| Product Name                            | Catalog Number | Indexes, Samples                 |
| --------------------------------------- | -------------- | -------------------------------- |
| Illumina RNA UD Indexes Set A, Ligation | 20091655       | <p>96 Indexes,<br>96 Samples</p> |
| Illumina RNA UD Indexes Set B, Ligation | 20091657       | <p>96 Indexes,<br>96 Samples</p> |
| Illumina RNA UD Indexes Set C, Ligation | 20091659       | <p>96 Indexes,<br>96 Samples</p> |
| Illumina RNA UD Indexes Set D, Ligation | 20091661       | <p>96 Indexes,<br>96 Samples</p> |

* Illumina RNA UD Indexes, Ligation are compatible with [Illumina Stranded Total RNA](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/stranded-total-rna-prep.html) and [Illumina Stranded mRNA](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/stranded-mrna-prep.html).
* Note that these kits contain an Illumina DNA/RNA UD Index plate as well as an RNA Index Anchor plate. The RNA Index Anchors are required for compatibility with the workflow.

**IDT for Illumina DNA/RNA Unique Dual Index, Tagmentation kits5**

| Product Name                                                        | Catalog Number | Indexes, Samples                 |
| ------------------------------------------------------------------- | -------------- | -------------------------------- |
| IDT for Illumina DNA/RNA UD Indexes Set A, Tagmentation<sup>1</sup> | 20027213       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina DNA/RNA UD Indexes Set B, Tagmentation<sup>2</sup> | 20027214       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina DNA/RNA UD Indexes Set C, Tagmentation             | 20042666       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina DNA/RNA UD Indexes Set D, Tagmentation             | 20042667       | <p>96 Indexes,<br>96 Samples</p> |

* All four index sets are compatible with [Nextera XT](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/nextera-xt-dna.html), [Illumina DNA Prep](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/nextera-dna-flex.html)3, and [Illumina DNA Prep with Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/nextera-flex-enrichment.html)4, [Illumina RNA Prep with Enrichment](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/rna-prep-enrichment.html) and [Illumina DNA PCR-Free](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/dna-pcr-free-prep.html).
* 1Formerly known as IDT for Illumina Nextera DNA UD Indexes Set A
* 2Formerly known as IDT for Illumina Nextera DNA UD Indexes Set B
* 3Formerly known as Nextera DNA Flex
* 4Formerly known as Nextera Flex for Enrichment

**IDT for Illumina RNA Unique Dual Indexes, Ligation kits5**

| Product Name                                    | Catalog Number | Indexes, Samples                 |
| ----------------------------------------------- | -------------- | -------------------------------- |
| IDT for Illumina RNA UD Indexes Set A, Ligation | 20040553       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina RNA UD Indexes Set B, Ligation | 20040554       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina RNA UD Indexes Set C, Ligation | 20040555       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina RNA UD Indexes Set D, Ligation | 20040556       | <p>96 Indexes,<br>96 Samples</p> |

* IDT for Illumina RNA UD Indexes, Ligation are compatible with [Illumina Stranded Total RNA](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/stranded-total-rna-prep.html) and [Illumina Stranded mRNA](https://www.illumina.com/products/by-type/sequencing-kits/library-prep-kits/stranded-mrna-prep.html).
* Note that these kits contain an IDT for Illumina DNA/RNA UD Index plate as well as an RNA Index Anchor plate. The RNA Index Anchors are required for compatibility with the workflow.
* 5Note: The IDT for Illumina DNA/RNA Unique Dual Index, Tagmentation and the IDT for Illumina RNA Unique Dual Indexes, Ligation kits are no longer available for purchase

**IDT for Illumina UMI DNA/RNA Unique Dual Indexes, Ligation kits**

| Product Name                                                    | Catalog Number | Indexes, Samples                 |
| --------------------------------------------------------------- | -------------- | -------------------------------- |
| IDT for Illumina UMI DNA/RNA UD Indexes Set A, Ligation         | 20034701       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina UMI DNA/RNA UD Indexes Set B, Ligation         | 20034702       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina UMI DNA/RNA Index Anchors Set A for Automation | 20066404       | <p>96 Indexes,<br>96 Samples</p> |
| IDT for Illumina UMI DNA/RNA Index Anchors Set B for Automation | 20063213       | <p>96 Indexes,<br>96 Samples</p> |

* IDT for Illumina UMI DNA/RNA UD Indexes are compatible with [TruSight Oncology 500 High-Throughput](https://www.illumina.com/products/by-type/clinical-research-products/trusight-oncology-500-ht.html).
* Note that these kits contain an IDT for Illumina DNA/RNA UD Index plate as well as an UMI DNA Index Anchor plate. The UMI DNA Index anchors are required for compatibility with the workflow.

**Ampliseq for Illumina Unique Dual Index kits**

| Product Name                     | Catalog Number | Indexes, Samples                 |
| -------------------------------- | -------------- | -------------------------------- |
| AmpliSeq UD Indexes for Illumina | 20019104       | <p>24 Indexes,<br>24 Samples</p> |

* AmpliSeq UDIs are compatible with all [AmpliSeq for Illumina](https://www.illumina.com/products/by-brand/ampliseq.html) library preparation kits

For a complete list of Illumina adapter sequences, refer to the [Illumina Adapter Sequences](https://support.illumina.com/downloads/illumina-adapter-sequences-document-1000000002694.html) document.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                              |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2344), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002344%20-%20Library%20Preparation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/library-preparation/general/library-preparation-general-reference_material-list/000002344.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.