For the complete documentation index, see llms.txt. This page is also available as Markdown.

Troubleshooting vacuum failures and flow cell vacuum is no longer engaged message on NovaSeq 6000

Vacuum issues can happen at the start of a run or during a run, and have different underlying causes.

1. Vacuum issues at the start of the run

2. Phenotype of runs with vacuum issues

  • When reviewing the run data in Sequencing Analysis Viewer (SAV), these manifest as low intensity and low Q30 at the inlet of the flow cell (see Figure 1).

  • This is caused by flow cell quality issues or glass shards at the bottom surface of the flow cell.

  • Glass shards on the flow cell stage may also impact flow cell vacuum engagement.

Figure 1: Flow Cell Chart displaying the detected intensity in SAV.

Troubleshooting steps:

  1. Inspect flow cell glass and flow cell stage.

  • If glass shards or debris are found, blot the flow cell glass or flow cell stage with an alcohol wipe to pick up the debris.

  • Once no debris is identified, clean the flow cell and flow cell stage with a lint-free alcohol wipe and dry with a low-lint lab tissue.

  • Check for reagent leakage on or around the flow cell holder.

  1. Check the flow cell vacuum pressure in NovaSeq Service Software.

  • Close the NovaSeq Control Software and launch the NovaSeq Service Software.

  • Once initialized, select Load Flow Cell.

  • Load wash flow cells on Sides A and B.

  • Expected Vacuum Pressure Values for wash flow cells are around -90 kPa and show a steady green check mark for the Vacuum Pressure State.

  • If the pressure is low or the Vacuum Pressure State is not steady, clean the flow cell and stage one more time (with an alcohol and a lint-free wipe) and re-check the Vacuum Pressure Values.

    • If the vacuum pressure is still outside the expected range, contact Illumina Technical Support (techsupport@illumina.com) with a screenshot of the flow cell vacuum pressure results.

  • If vacuum pressure is within the expected range, perform a power cycle and proceed with the next run.

For any feedback or questions regarding this article (Illumina Knowledge Article #1309), contact Illumina Technical Support techsupport@illumina.com.

Last updated

Was this helpful?