> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-reference_material-list/000002004.md).

# Thawing and Stability Guidelines for MiniSeq Reagent Kits

See the following for thawing and stability guidelines for MiniSeq reagent cartridges, flow cells, and hybridization buffer.

**MiniSeq Reagent Cartridge (store at -20°C)**

* Thawing options:
  * Thaw at 37°C in a water bath for 35 minutes; the cartridge is stable for up to 2 hours.
  * Thaw at Room Temperature water bath (19°C to 25°C) for 90 minutes; the cartridge is stable for up to 24 hours if moved to 4°C.
  * Thaw at room temperature in air (no water bath) for 5 hours; the cartridge is stable for up to 24 hours if moved to 4°C.
  * Thaw at 2°C to 8°C (fridge) for 18 hours; the cartridge is stable for up to 72 hours at 4°C.
* If reagents need to be stored during troubleshooting, store them at 4°C until ready to run within the stability time frames listed above.

**MiniSeq flow cell (shipped dry, store 2-8°C until use)**

* Let sit for 15-30 minutes at room temperature prior to opening the package.
* The flow cell can be returned to 4°C **once** if unopened.
* Use flow cell within 12 hours once opened.
  * Flow cells are sensitive to moisture; clustering can be negatively impacted if the flow cell package is left open for too long prior to starting a run.

**Hybridization Buffer (5 ml; store at -20°C)**

* Thaw at room temperature and store on ice until ready for use.

More information, please reference [MiniSeq System Guide](https://support.illumina.com/content/dam/illumina-support/documents/documentation/system_documentation/miniseq/miniseq-denature-dilute-guide-1000000002697-09.pdf).

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #2004), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000002004%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/miniseq/instrumentation-miniseq-reference_material-list/000002004.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.