> For the complete documentation index, see [llms.txt](https://knowledge.illumina.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007553.md).

# How to install TeamViewer QuickSupport on the NovaSeq X Series Instruments

Background

The TeamViewer QuickSupport application allows users to provide Illumina Support teams with remote access to the NovaSeq X Series instruments for troubleshooting assistance. The following steps can be used to install the QuickSupport application to the instrument when needed.

**IMPORTANT NOTE:** The QuickSupport software is safe to use during instrument operation, though it should be closed whenever not in use to allow the NovaSeq X Series instrument access to all available computer resources.

Use the following steps to install TeamViewer QuickSupport on the NovaSeq X Series.

1. On the instrument, open the pre-installed Chromium browser and navigate to share.illumina.com.
2. Under the **Using TeamViewer** header, select the **Linux** link.
3. Once the TeamViewer web page loads, scroll down and then select the **Download QuickSupport** button.
4. Once the download finishes, locate the **teamviewer\_qs.tar.gz** download status in the bottom-left corner of Chromium, then select the ^ button and select **Show in folder** (Figure 1).
5. Once the Files application opens to the default Downloads directory, right click on **teamviewer\_qs.tar.gz** and select **Extract Here** (Figure 2).
6. Once the archived file finishes extracting, navigate to **Home > Downloads > teamviewer\_qs > teamviewerqs** in the Files application (Figure 3.1).
7. Double-click the **teamviewer** application to start TeamViewer QuickSupport (Figure 3.2).

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-0e25a9c6b5c40932c874188e86f73c50ee267e13%2Fimage1.png?alt=media\&token=4ee538f4-7f76-416f-b839-5595a4e3fe63)

Figure 1: Selecting **Show in Folder** in Chromium after teamviewerqs.tar.gz download completes.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-5c213236e8cf3c621505b9a55d43f4bce9f8718b%2Fimage2.jpg?alt=media\&token=eb129bb8-71de-4402-9ea8-2eace45b248c)

Figure 2: Extracting the TeamViewer QuickSupport application from the teamviewer\_qs.tar.gz archived file.

![](https://761066130-files.gitbook.io/~/files/v0/b/gitbook-x-prod.appspot.com/o/spaces%2FGM9W2DuBTgEXv1ClCm8H%2Fuploads%2Fgit-blob-fb3e482fd7fd244b8c960d912cd0f2543c342afd%2Fimage3.png?alt=media\&token=b2e865de-e336-4162-8953-3997858c7501)

Figure 3: 1) Directory path to the TeamViewer QuickSupport application. 2) The TeamViewer QuickSupport application within the teamviewerqs directory.

\
\
\ <br>

|                                                                                                                                                                                                                                                                                                                                                                        |
| :--------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------------: |
| *For any feedback or questions regarding this article (Illumina Knowledge Article #7553), contact Illumina Technical Support* [*techsupport@illumina.com*](mailto:techsupport@illumina.com?subject=Question%2FFeedback%20Regarding%20Illumina%20Knowledge%20Article%20#000007553%20-%20Instrumentation%20\&body=Dear%20Illumina%20Technical%20Support,%0D%0A%0D%0A)*.* |


---

# Agent Instructions
This documentation is published with GitBook. GitBook is the documentation platform designed so that both humans and AI agents can read, navigate, and reason over technical content effectively. Learn more at gitbook.com.

## Querying This Documentation
If you need additional information that is not directly available in this page, you can query the documentation by asking a question.

Perform an HTTP GET request on the following URL with the `ask` and `goal` query parameters:

```
GET https://knowledge.illumina.com/instrumentation/novaseq-x-x-plus/instrumentation-novaseq-x-x-plus-reference_material-list/000007553.md?ask=<question>&goal=<user_goal>
```

`ask` is the immediate question: it should be specific, self-contained, and written in natural language.
`goal` is what the user is ultimately trying to achieve, the reason they need the answer. Sharing it helps GitBook give you a better, more relevant answer. A goal is most helpful when it describes the outcome the user wants rather than restating the question. For example, with `ask=how do I create an API token`, a goal like `automate deployments from our CI pipeline` lets GitBook tailor the answer to that use case.

The response will contain a direct answer to the question and relevant excerpts and sources from the documentation.

Use this mechanism when the answer is not explicitly present in the current page, you need clarification or additional context, or you want to retrieve related documentation sections.